background image

4. SETTING UP YOUR BLOOD PRESSURE MONITOR

4.1 Inserting the batteries

After you have unpacked your device, insert the batteries. The battery 
compartment is located on the back side of the device (see illustration).  
a) Remove cover from the bottom plate, as illustrated.
b) Insert the batteries (4 x size AA ). Always use AA long life batteries 

or alkaline 1.5v batteries. Do not use rechargeable batteries.

c) If a battery warning appears in the display, the batteries are discharged 

and must be replaced.

Attention!

• After the battery warning appears, the device will not operate until the batteries have been replaced. 
• If the blood pressure monitor will be left unused for long periods, remove the batteries from the device.

Functional check: Hold the ON/OFF (START) button down to test all the display elements.                  
When functioning correctly all segments must appear.

4.2. Using an AC/DC power adapter (special accessory)

It is possible to operate this blood pressure instrument with an AC/DC adapter. (output 6 V DC / 600 mA 
with DIN plug). Make certain that you use an adapter which fulfills the legal requirements and electronic 
requirements in the U.S. (UL standard)

a) Insert the plug into the left socket on the left side of 

the instrument, as shown in the diagram.

b) Plug the AC adapter into a 110 V

power socket (U.S. or Canada).

c) Test that power is available by                         

pressing the ON/OFF button.

Note:

• No power is taken from the batteries while the AC/DC adapter is connected to the instrument.
• If the power is interrupted during a measurement (e.g. by removal of the adapter from the wall socket), 

the instrument must be reset  by removing the plug from the instrument.

• Consult ADC

®

if you have questions

relating to the AC/DC adapter.

Warning:

• Use only ADC

®

or ADC

®

approved adaptors on your

blood pressure unit. ADC

®

is not responsible for

damages caused by the use of non-ADC

®

adaptors.

Using a non-ADC

®

adaptor voids the warranty on your

digital blood pressure unit.

4.3. Tube connection

Insert the cuff tube into the right socket 
on the left side of the instrument, 
as shown in the diagram. 

Right Opening

Left Opening

7

12

5.5. Discontinuing a measurement

If it is necessary to interrupt a blood pressure 
measurement for any reason (e.g the patient feels 
unwell), the START button can be pressed at any time. 
The device then immediately lowers the cuff pressure 
automatically.

5.6. Memory – displaying the last 14 measurements

The measured results are stored in the instrument until a new 
measurement is carried out or the batteries are removed. With the
unit OFF, press and hold the MEMORY button for at least
3-seconds. The screen will show the last reading along with time and date. 
The MR symbol is displayed in the lower left. Release the button.The last 
reading will be displayed for 3 more seconds. Each time you press the
memory button an earlier measurement will be displayed along with time and date.  

Further information

Measurements should not occur soon after each other, since 
the results will be inaccurate. Wait 5 minutes in a relaxed position, 
sitting or lying down, before you repeat a measurement.

6. ERROR MESSAGES/TROUBLESHOOTING

If an error occurs during a measurement, the measurement is 
discontinued and a corresponding error code is displayed.
(example: Err 1)

Error No.

Possible cause(s) / Solutions

Err 1

The systolic pressure was not determined. The tube
may have loosened, or no pulse was detected.
*Ensure cuff connections are tight with 
proper cuff placement. See Section 5.3.

Err 2

Unnatural pressure impulses.
The arm was moved during the measurement. 
Repeat measurement keeping arm still.

MR

MR

Last
reading

14 readings
in memory

14

TIME           9:17

Summary of Contents for ADvantage 6013

Page 1: ...Digital Blood Pressure Monitor Automatic Blood Pressure Monitor 6013 Instruction Manual ADC 55 Commerce Drive Hauppauge NY 11788 ADC UK Ltd Unit 6 PO14 1TH United Kingdom EC REP Inspected in the U S A...

Page 2: ...Record My Blood Pressure 3 Components of Your Blood Pressure Monitor 4 Setting Up Your Blood Pressure Monitor 4 1 Inserting the Batteries 4 2 Using an AC Power Adapter 4 3 Tube Connection 4 4 Setting...

Page 3: ...hypertension Under no circumstances should YOU alter the dosages of any drugs prescribed by your doctor This monitor is intended for use by adults only Consult with a physician before using this inst...

Page 4: ...and react appropriately If you are undergoing medical treatment to control your blood pressure keep a record of values along with time of day and date Show these values to your doctor Never use the r...

Page 5: ...physician is qualified to interpret the readings obtained from your blood pressure monitor No attempt should ever be made at self diagnosis or treatment 15 4 Range Systolic Diastolic Precaution Classi...

Page 6: ...of the remaining readings 2 8 How is Blood Pressure Measured Health care professionals traditionally use a device known as a sphygmomanometer along with a stethoscope essentially a professional versio...

Page 7: ...on Pulse Cuff Holder 13 Repeat measurement keeping arm still If inflation of the cuff takes too long the cuff is not correctly seated or the tube connection s not tight Check connections and repeat Th...

Page 8: ...g Use only ADC or ADC approved adaptors on your blood pressure unit ADC is not responsible for damages caused by the use of non ADC adaptors Using a non ADC adaptor voids the warranty on your digital...

Page 9: ...isplay now switches to the current date during which the first character month blinks Note Holding the button down speeds up the procedure 4 The current month can now be entered by pressing the MEMORY...

Page 10: ...inute rest to ensure accuracy 5 3 Fitting the cuff a Position the cuff flat on a table with the hook and loop adhesive side down Pass the end of the cuff through the flat metal ring so that a loop is...

Reviews: