VA4runup - runup projects are exported here
VA4recorder - records are exported here
va4ver file - unit version information (for DDS)
These folders are created when the unit needs them, not actually in the first run.
A4410 Virtual unit with the DDS connection
Press the Find button in the DDS VA4Pro window for instrument connection and find manually (Find button)
A4410_VirtualUnit_verXXXX\VA4_DISC virtual unit folder or the A4410_VirtualUnit_verXXXX\Data folder.
Connection of the Virtual unit with the DDS is done the same way as with the A4400 Va4Pro instrument. It is
possible to copy projects to and from DDS to and from the A4410 Virtual Unit.
If the DDS and the Virtual unit are not installed on the same computer then you do not reach the DDS via
network. It is possible to copy the VA4_DISC to the flash drive memory which can be connected to the computer
with DDS. For DDS is important to find the standard VA4_DISC folder structure. It is not necessary to copy all
VA4_DISC folders, however directory structure has to be kept (that means the Analyzer projects has to be in
VA4analyser folder, etc.).
!!! Attention: VA4_DISC which should be read by DDS have to contain va4ver file. Don’t forget to copy this file,
when you copy the VA4_DISC folder.
A4410 Virtual unit and A4400 VA4Pro connection
How to share projects between the real and virtual unit will be described in this chapter.
Connect the real Va4Pro instrument to the computer with installed Virtual unit (via USB cable). The VA4Pro unit
will be found as VA4_DISC (removable disc) on your system. It’s possible to copy the data to and from the
Virtual VA4_DISC folder.
Copying projects to the VA4Pro instrument
Copy the A4410_VirtualUnit_verXXXX\Data\VA4analyser\ProjectName project folder (ProjectName is your
folder name which you want to copy) into the VA4Pro VA4_DISC\VA4analyser folder (VA4_DISC is the real
removable storage device here, NOT the virtual unit folder !!!). You have to create empty file with the name
script.dds in VA4Pro VA4_DISC\VA4analyser after project copy (it has the label sense only). RunUp and
Route projects (VA4runup and VA4route folders) are copied the same way. Next mode projects are not
available to copy.
Example:
We’ve got the analyzer project called Power Station in the virtual unit and we want to add it to the real
instrument.
1) Connect the instrument with the computer.
2) VA4_DISC removable storage device appears. Open VA4analyser folder on this device.
3) Open \A4410_VirtualUnit_verXXXX\Data\Va4analyser folder. Find Power Station folder in this folder. Copy
folder into the VA4Pro VA4_DISC.
4) Create the empty file script.dds in VA4Pro VA4_DISC\VA4analyser\Power Station folder.
Projects and records copy from the VA4Pro instrument
!!! Attention: Projects has to be exported to the VA4Pro VA4_DISC first (the same as export to the DDS).
Copy the VA4Pro VA4_DISC\VA4analyser\ProjectName (the real removable storage device here, NOT the
virtual unit folder!!! ) into the A4410_VirtualUnit_verXXXX\Data\VA4analyser folder. In this case there is no