background image

For detailed instructions on how to use the software, please visit

PROTON WHO

Nature of Violations and Exposure intervals are shown in below table:

User Interface

User can fetch the data from PC/Laptop, also can see data on display by clicking on prompt button.

User can see number of violations, type of violation, time stamps of violations, trip time on the display.

Time stamps can be seen by clicking on the button in steps.

If there are No violations,  Tick/OK mark will be displayed on the LCD.

If there are violations, X/not OK mark will be displayed on the LCD.

www.adaptloggers.com

+91-8639739890

[email protected]

https://docs.adaptloggers.com/devices/proton/protonswho/

03

END

hrs

Trip Ended with Violations

Higher Threshold Violation

Recording Ended

Medium Threshold Violation

Lower Threshold Violation

Violation Count

END

hrs

Trip Ended with no violations

Recording Ended

OK

OK

Alarm settings

 

  

A

larm T

y

pe

 

Time

-

Temperature 

 

threshold

 

P

eriod of 

E

xposure

 

T

y

pe 

C

 

High Threshold

 

45

˚C

 single e

v

ent

 

1

 Hours

 

Medium Threshold

 

≥30˚C

 cumulati

v

e exposure 

 

10

 Hours

 

Low Threshold

 

-

0.

5

˚C

 single e

v

ent

 

1

 Hours

 

Type A/B 

High Threshold

 

45

˚C

 single e

v

ent

 

1

 Hours

 

Medium Threshold

 

≥30˚C

 cumulati

v

e exposure 

 

10

 Hours

 

Low Threshold

 

≥10˚C

 cumulati

v

e exposure

 

2

0

 Hours

 

Type 

Rotateq 

High Threshold

 

2

7˚C

 single e

v

ent

 

1

 Hours

 

Medium Threshold

 

≥17˚C

 cumulati

v

e exposure

 

2 Hours

 

Low Threshold

 

 

-

25

˚C

 single e

v

ent

 

1

  Hours

 

Type 

Prevenar 

High Threshold

 

4

0˚C

 single e

v

ent

 

1

 Hours

 

Medium Threshold

 

≥30˚C

 cumulati

v

e exposure 

 

10

 Hours

 

Low Threshold

 

-

0.

5

˚C

 single e

v

ent

 

1

 Hours

 

 

Reviews: