© Adam Equipment Company 2021
22
10.4.4.3
Animal (Dynamic) Weighing Setup Parameters
When the “
Animal
” text is displayed, and you have selected
“
ΣΕτϖΠ
”
to set up the balance
for dynamic weighing (follow the set up instructions below):
The display will show
“
ΜΟ∆Ε
”.
Press [Setup] again and use the [Up] or [Down] keys to
select
“
ΑϖτΟ
”
or
“
µµΑνϖΑΛ
”.
If
“
ΑϖτΟ
”
[deleted text]
is selected, the following 3 parameters are available
:
•
Threshold
“
ΤΗΡΕΣΗ
”
•
Test time
“
ΤΕΣΤ Τ
”
•
Delay
“
∆ΕΛΑΨ
”
If “
µµΑνϖΑΛ
” is selected,
the following 2 parameters are available:
•
Test time
(“
ΤΕΣΤ Τ
”)
•
Delay
“
∆ΕΛΑΨ
”
10.4.4.3.1 Threshold
“
ΤΗΡΕΣΗ
” (For Auto mode only)
Press [Setup] when
“
ΤΗΡΕΣΗ
”
is shown and the display will next show the minimum weight
of the item required by the balance to start the process for dynamic weighing. The value
shown will be the current value in the last selected unit.
The minimum threshold value can be changed from 1.0 to 100 grams using the keypad
numeric entry method.
To confirm the desired value, press [Setup] or to escape without changing the value, press
[Mode].
10.4.4.3.2
Test time
“
ΤΕΣΤ Τ
”
Press [Setup] when
“
ΤΕΣτ τ
”
is shown and the display will next show the number of
seconds over which the balance will average to compute the final weight. The Test time
value can be changed to between 10 - 99 seconds using the keypad numeric entry method.
To confirm the desired value, press [Setup] or to escape without changing the value, press
[Mode].
10.4.4.3.3
Delay
“
∆ΕΛΑΨ
”
Press [Setup] when
“
∆ΕΛΑΨ
”
is shown and the display will next show the number of
seconds pause before the sampling starts. The Delay value can be changed to between 0-
99 seconds using the keypad numeric entry method.
To confirm the desired value, press [Setup] or to escape without changing the value, press
[Mode].
10.4.4.3.4
Mode
“
µοδε
”
Auto
“
ΑϖτΟ
”
or Manual
“
µµΑνϖΑΛ
”
modes are available. Whichever mode is visible when
[Setup] is pressed becomes the active mode. Auto starts dynamic weighing test as soon
as weight exceeding a set threshold is loaded on the pan. Manual requires the user to load
the pan and then press a button before weighing commences
.