Adam Equipment Company 2020
Add another sample to the top pan and press the
[M+]
key; the memory entry will
be incremented by 1, the middle display will show total weight of all samples
counted in memory and the right display will show the total count of all samples
in memory.
To recall the total of the memory function, press the
[M+]
key when there is no
weight on the top pan.
To clear the memory, press
[M+]
key when there is no weight on the top pan and
then press the
[CE]
key; the display will briefly show ACC 0 to indicate the
memory has been cleared.
6.0 CALIBRATION
Calibration before use is recommended for best accuracy. It is suggested that
the scale be calibrated when first installed and any time the scale is moved to a
different environment.
Ensure weights are ready before calibrating, the weight should be of a known
accuracy, for example OIML class M1 or Class 4 of ASTM E617.
Do not calibrate the scale if a proper calibration weight is not available.
6.1 USER CALIBRATION PARAMETER SETTING
When power on, press TARE during self-checking, enter pass code 1000 to access user
calibration parameter.
Use keys 1 and 6 to select available options and press TARE button to conform the
selection.
U-CAL (user calibration)
This parameter sets the calibration of the scale. Press [TARE]; display will show unload;
with no weight on the top pan and the stable indicator On, press [TARE] key to set the
zero point. The display will then indicate the load (SET) amount. Use Keys 1 and 6 and
CE to enter the desired load amount and then press the [TARE] key. The display will then
show the pre-selected (LOAD) amount for calibration; put this weight on the top pan, wait
for the stable indicator to show on the display and then press the [TARE] key.
AdCnt
– this will show the internal A/D counts.