© Adam Equipment Company 2014
19
9520revF-Jun14
and press
[Print]
to decrease the value. The internal calibration weight should
only change between
95.000
and
105.000
grams or
495.00
and
505.00
grams
(depending upon the model).
6)
Press
[Tare]
to store a particular setting. The display will return to
F6
CA
7)
When
F6 CA
is displayed press the
[Print]
key to return to weighing or press
[Mode]
to go to the next function.
Changing the value of the internal weight will affect the calibration when using the
internal calibration. To check if your internal weight is correct you should only use
high quality weights that are accurate to the readability of the balance.
6.7
SETTING
USER
PARAMETERS
The balance has a number of parameters the user can adjust.
Display Description
Default Value
n FIL
Filter settings, 1, 2 or 3
1 is the fastest filter, 3 is the slowest.
1 FIL
n ZEO
The Automatic Rezero range
0 to 8, 0 being with auto zero off, and 8 is the zero
range of 4 divisions.
4 ZEO
n STA
Stability symbol range, 0 to 5
0 is the smallest stability range and 5 is largest.
2 STA
n ZTR Weight Tracking Range, 0 to 5
3 ZTR
Press the
[Tare]
key to view the current settings.
To change the settings press the
[Mode]
key. Press the
[Tare]
key to advance to
the next parameter. When completed the display will show
F7 SET
again.
Summary of Contents for HIGHLAND HCB1002
Page 2: ... Adam Equipment Company 2014 ENGLISH P1 30 FRANÇAIS P31 59 DEUTSCH P60 90 ESPAÑOL P91 116 ...
Page 89: ... Adam Equipment Company 2014 87 9520revF Jun14 10 0 KALIBRIER ZERTIFIKAT ...
Page 115: ... Adam Equipment Company 2014 113 9520revF Jun14 10 0 CERTIFICADO DE CALIBRACIÓN ...
Page 119: ... Adam Equipment Company 2014 ...