© Adam Equipment Company 2007
17
if 3500g is placed on the scale and percent weighing is selected,
the
display will show 100.00%. Remove the 3500g weight and place a
3000g weight. The display will show 85.7% as 3000g is 85.7% of
3500g.
•
The number of decimal points will depend on the weight used. A
smaller weight will show only “100%” while a larger weight might show
“100.00%”.
•
If the scale was showing zero weight when entering this function, then
the user must manually enter the weight to be set as 100%. Press
[Tare]
to accept the reference weight. The display will show “
0.00 %
”.
•
If the scale shows “
x x . x x %
”, which is the last weight used as a
reference, press
[CE]
to clear and use the numeric keypad to enter a
new value. Press
[Tare]
to accept the new reference weight.
•
The weight entered must be greater than 50 scale divisions.
•
Press
[Unit]
to return to weighing.
NOTE:
The display may jump by large numbers unexpectedly if small weights are
used to set as 100% reference. For example, if only 23.5g is placed on a
scale with 0.5g increments and is set to 100%, the display will show
100.00%. However a small change of weight will cause the display to jump to
102.13% as an increase of one scale division (0.5g) to 24.0g will be
equivalent to an increase of 2.13%.
10.8 ACCUMULATED TOTAL
The scale can be set to accumulate when a weight is added to the scale
automatically or manually by pressing
[Print]
. See Section 13.2.