@Adam Equipment Company 2007

 

 

 

3

2.0 

TROUBLE SHOOTING GUIDE  

 
PROBLEM   

 

 

POSSIBLE CAUSE

 

Display is blank 

Power Switch is faulty or not turned on 

No turn on test 

Battery is faulty or not charged 

 

Power supply is incorrect 

 
Display all zeros after turn on 

Pan not installed 

or E6 is displayed 

Unstable weight 

 

Load cell damaged 

 Electronics 

faulty 

 

  

‘- - - - -‘ appears on display 

Maximum capacity exceeded 

 

Load Cell damaged 

 

Electronics is faulty 

 
Display is unstable 

Drafts or air currents 

 

Obstruction under pan 

 

Sample is moving (animal weighing) 

 

Vibrations through table 

 

Temperature changed dramatically 

 Electronics 

faulty 

 Battery 

low 

 

 

Weight value incorrect 

Calibration error, Recalibrate 

 

Unit calibrated with inaccurate weight 

 

Balance not level 

 

Obstruction between sample and cover  

 

Wrong unit of weight displayed 

 
Cannot use Full Capacity   

Over-load Stops hitting pan support or hitting    

     bottom 

of 

load 

cell 

     Electronics 

faulty 

 

     Parameters 

set 

incorrectly 

     Load 

Cell 

Damaged 

 
Not Linear 

 

 

 

Overload stops hitting too soon 

     Faulty 

Factory 

Calibration 

 
Off Centre Loading error   

Overload Stops not correct 

     Obstruction 

under 

pan 

 
Battery will not charge 

 

Charging circuit failure 

     Battery 

failure 

     Incorrect 

adapter 

(below 

800 

mA) 

 

 

 

 

 

Processor failure

Summary of Contents for CBD Series

Page 1: ...Adam Equipment CBD SERIES SERVICE MANUAL ADAM EQUIPMENT CO LTD P N 7951 Rev A1 March 2007 Adam Equipment Company 2007...

Page 2: ...Adam Equipment Company 2007 2...

Page 3: ...0 TROUBLE SHOOTING GUIDE 3 3 0 ERROR CODES 4 4 0 CBD DESCRIPTION 4 5 0 SETTING UP THE REMOTE SCALE 5 6 0 USER PARAMETERS 7 7 0 USER CALIBRATION 9 8 0 TECHNICAL PARAMETERS 10 9 0 LOAD CELL DAMAGE 12 10...

Page 4: ...e Contact telephone e mail fax or any other methods B Details of the unit purchased This part of information should always be available for any future correspondence We suggest you to fill in this for...

Page 5: ...ng animal weighing Vibrations through table Temperature changed dramatically Electronics faulty Battery low Weight value incorrect Calibration error Recalibrate Unit calibrated with inaccurate weight...

Page 6: ...of the scale Damaged load cell Damaged Electronics Err 5 Keyboard error Improper operation of the scale Err 6 A D count is not correct when turning the scale on Platform is not installed Load cell may...

Page 7: ...assembly includes 9 A D converter AU1 9 Microprocessor DU1 and 9 Power supply battery charger circuits DU5 T2 5 0 SETTING UP THE REMOTE SCALE The CBD Series can be connected to any size of load cell...

Page 8: ...tion with a maximum of 1 30 000 i e 300kg x 10g It will also be possible to sample on the remote scale with the same accuracy as the Local Where more than one load cell is fitted or the total load cel...

Page 9: ...tting press the Print key To change and scroll through the other available settings press U Wt To choose the desired setting and thereby go back to the sub parameter press the Print key To return to t...

Page 10: ...speed for the RS 232 communications Default rate is 4800 F2 Prt PAritY 8 n 1 7 E 1 7 o 1 8 data bits no parity 7 data bits even parity 7 data bits odd parity U id U id Abc234 Shows the current user ID...

Page 11: ...h LocAL tEch rEmo tE Use the U Wt to select the weighing unit to be used for setting of the calibration The arrow in the Weight window will indicate the unit selected Press the Print key to continue t...

Page 12: ...9 will override any other user password Press Print Pi n Select the scale to be set up by using the Local Rem key Press the Print to enter the technical section for that scale t Ech Lo c A L t Ech r E...

Page 13: ...t Unit to increment the values then Print Azt Autozero tracking range Select from 0 5d 1d 2 4 d Press the U Wt Unit to go to next menu use Print key to enter Use U Wt Unit to increment the values then...

Page 14: ...ges to consider Normally without a load on the load cell all the resistances are the same However when the load cell has a weight on it two of the strain gauges will be in compression A and D and the...

Page 15: ...d cell is connected to the power supply use the ohmmeter to measure the resistance between the wires Typical resistance values are E to E 410 ohms 30 ohms S to S 350 ohms 2 ohms E to S or any other co...

Page 16: ...cell has been damaged the no load voltage will likely be greater than 3mv or the loaded voltage will be grossly different from the expected value Remember the loaded voltage will be offset by the amou...

Page 17: ...In use cover RS 232 option Printer etc 11 0 SCHEMATICS The following schematics are included as a guide for the experienced Service Technician If you are not experienced we do not advice you to chang...

Page 18: ...ell 1 0 0008 12 28 Screw M4 x 12 4 1 0047 13 Washer 4 1 0048 14 Main PCB 1 1 0022 15 Display PCB 1 0011 16 Bottom Cover 1 0 0011 17 Case screw M4 x 15 5 1 0049 18 Serial Port RS 232 1 1 0050 19 Duel S...

Page 19: ...Adam Equipment Company 2007 17...

Page 20: ...ill cease to operate if the equipment is not returned in the original packaging and with correct documentation for a claim to be processed All claims are at the sole discretion of Adam Equipment This...

Page 21: ...y to cause harmful interference in which case the user will be required to correct the interference at his own expense Shielded interconnect cables must be employed with this equipment to insure compl...

Page 22: ...ns without the prior permission of Adam Equipment Adam Equipment reserves the right to make changes to the technology features specifications and design of the equipment without notice All information...

Reviews: