@Adam Equipment Company 2007
4
3.0 ERROR
CODES
ERROR CODE
DESCRIPTION
POSSIBLE CAUSES
Err 4
Initial Zero is greater than
the permissible value
(typically 4% of maximum
capacity) when power is
turned on or when the
[Zero]
key is pressed,
Weight on the pan when
turning the scale on.
Excessive weight on the pan
when zeroing the scale.
Improper calibration of the
scale.
Damaged load cell.
Damaged Electronics.
Err 5
Keyboard error.
Improper operation of the
scale.
Err 6
A/D count is not correct
when turning the scale on.
Platform is not installed.
Load cell may be damaged.
Electronics may be damaged.
FAIL H
or
FAIL L
Calibration error
Improper calibration.
If the problem persists
contact your dealer or Adam
Equipment for assistance.
4.0 CBD
DESCRIPTION
The CBD scales have an enclosure with all components mounted within it. To
gain access to the components remove the 4 screws securing the cover to the
base.