3.0  ERROR CODES 

 

ERROR CODE  DESCRIPTION 

POSSIBLE CAUSES 

Initial Zero is  greater than  allowed 

(typically  4%  of  the  maximum 

capacity) when power is turned on 

or when the 

[Zero]

 key is pressed, 

Weight  is  on  the  pan  when 

turning the scale on. 

 

Excessive  weight  on  the  pan 

when zeroing the scale. 

 

Improper  calibration  of  the 

scale. 

 

Damaged load cell. 

 

Damaged Electronics. 

 

Keyboard error.   

Improper operation of the scale. 

 

A/D  count  is  not  correct  when 

turning the scale on. 

Platform is not installed. 

 

Load cell is damaged. 

 

Electronics is damaged. 

 

or 

or 

 for 

some softwares

Calibration error. 

Improper  calibration  (should  be 

within  +  10%  of  the  factory 

calibration).  The  old  calibration 

data  will  be  retained  until  the 

calibration process is complete. 

 

Scale is unstable. 

There  is  vibration  or  draft 

making the scale unstable. 

 

Electronics may be damaged. 

 

 

 

 

 

Summary of Contents for CBC 15

Page 1: ...Adam Equipment Company 2005 Adam Equipment CBC SERIES SERVICE MANUAL ADAM EQUIPMENT CO LTD p n 4329 Rev A3 January 2005...

Page 2: ...Adam Equipment Company 2005 2...

Page 3: ...6 4 0 CBC DESCRIPTION 7 5 0 CBC PARAMETERS AND CALIBRATION VER 1 2 2 02 8 5 1 CBC PARAMETERS VER 1 2 2 02 8 5 2 CBC CALIBRATION VER 1 2 2 02 9 6 0 FACTORY PARAMETERS VERSION 1 1 1 8 TO 1 2 2 02 10 7 0...

Page 4: ...your dealer or Adam Equipment for more assistance In order to supply further assistance we will need the following information Name of your company Contact Name Contact telephone e mail fax or other m...

Page 5: ...ng animal weighing Vibrations through table Temperature changed dramatically Electronics faulty Battery low Weight value incorrect Calibration error Recalibrate Unit calibrated with inaccurate weight...

Page 6: ...alibration of the scale Damaged load cell Damaged Electronics Keyboard error Improper operation of the scale A D count is not correct when turning the scale on Platform is not installed Load cell is d...

Page 7: ...C are similar except the selection of load cells and the program The main circuit board is the same as used in the CBW scales Normally if a problem is found with a circuit board the most cost effectiv...

Page 8: ...the scale increment Press the U Wt Units key to cycle through the options 1 2 5 10 Press Tare to accept the selection Displays the A D counts Press Tare to return to menu Press the U Wt Units key to...

Page 9: ...t on the platform after the stable symbol is on or enter the value desired and then press Tare The display will show Place the weight on the platform and press Tare again At all times the scale should...

Page 10: ...nting of the display after power turned on The Weight Display will show requesting the password number to be entered The factory password is and will over ride any other password they may have been pr...

Page 11: ...is that the load cell has been damaged The damage can be from 2 primary causes The first is physical damage due to an overload or an impact from the side and the second is damage due to the environmen...

Page 12: ...al output at S and S will show a voltage of zero volts with no load and a small voltage proportional to the load as the load is increased Typical load cells show 20 30mv of signal if the excitation vo...

Page 13: ...or 3mv V The expected change to the output is Vout 2mv V Vext Mass capacity of load cell For example a 2mv V load cell of 30Kg capacity with 20Kg placed on it Vext is 5 volts would give a output of a...

Page 14: ...arameters and Calibration including the Two Point Factory Calibration must be set before the new PCB will be put to use for weighing The 3 pin connector from the RS 232 connecter goes to the main PCB...

Page 15: ...processor and the links in the same way as it was in the old board into the new board Be very careful not to bend the legs while removing the micro processor Plug the cables coming from the Load cell...

Page 16: ...t advice you to change parts on the circuit boards or otherwise disturb the electronics If you require assistance please contact your dealer or Adam Equipment We will be happy to help you resolve prob...

Page 17: ...Adam Equipment Company 2005 17...

Page 18: ...Adam Equipment Company 2005 18...

Page 19: ...to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment The equipment generates uses and can radiate radio frequency energy and if not...

Page 20: ...tronic Scales for Medical use Retail Scales for price computing Adam Equipment Co Ltd Bond Avenue Milton Keynes MK1 1SW UK Phone 44 0 1908 274545 Fax 44 0 1908 641339 e mail sales adamequipment co uk...

Reviews: