The parameters available are:
F1
CAL
To enter the Calibration
F2
rES
Resolution
selection
F3
Unit
Select kilograms or pounds
F4
CAPA
Select
capacity
F5
Point
Unit Price/Price decimal point
F6
init ZEro
Initial Zero Range
F7
rE ZEro
Re-Zero range
F8
SCSivE tArE
Successive Tare Enable
F9
Ad CoUnt
Display the A/D counts
F10
roUnd
Rounding off the Total Price
F11
ZEro
Select the zero mode
F12 Lvd
modE
Low voltage detection
6.2.1
F1
-CALIBRATION
To enter the calibration parameter, press the
[Tare]
key when “
F1
” is displayed.
The display will instruct you to remove any weight from the scale, “
UnloAd
”. Press
[Tare]
.
The display will tell you to add weight to the scale: “
LoAd
” “
15
” “
kilo
”
Add the weight shown, wait for stability then press the
[Tare]
key.
The display will show “
SPAn
” “
PASS
” if the calibration is OK. Or it will show “
SPAn
” “
FAIL
” if
calibration could not be completed correctly.
The display will then show “
JP
” “
On
” showing the jumper is still in place. Remove the
weight. Switch off the scale and switch it on again to continue with the other Service
parameters.
Summary of Contents for AZextra 15
Page 2: ...Adam Equipment Company 2007...
Page 20: ...7 0 WIRING DIAGRAM Adam Equipment Company 2007 18...
Page 21: ...8 0 MECHANICAL ASSEMBLY Adam Equipment Company 2007 19...
Page 23: ...9 0 PILLAR ASSEMBLY Adam Equipment Company 2007 21...
Page 24: ...10 0 FRONT DISPLAY ASSEMBLY Adam Equipment Company 2007 22...
Page 25: ...11 0 REAR DISPLAY ASSEMBLY Adam Equipment Company 2007 23...
Page 26: ...12 0 PILLAR DISPLAY ASSEMBLY Adam Equipment Company 2007 24...