@Adam Equipment Company 2005
16
9.1 LINEARITY
Press
[Zero]
when “
#
” is displayed.
Next “
"
” will be displayed.
Press
[Print],
[Smpl]
and the
[%]
key in this order and press
[Zero]
.
Display will show “
”. Press
[Zero]
“1/3 LoAd”, Press [Zero] “3/3 LoAd”
Press
[Zero].
The balance will return to normal weighing.
9.2 FACTORY CALIBRATION
Press
[Zero]
when “
” is displayed.
See section 10.0 for details on Factory Calibration.
9.3 A/D COUNT
Press
[Zero]
when “
” is displayed.
The A/D Count will be displayed. If the Load cell is good then Counts should operate
between 10,000 and 900,000 with maximum load.
Press
[Zero]
to escape this parameter.
Press
[Mode]
to go to the next technical parameter.
9.4 AUTO ZERO RANGE
Press
[Zero]
when “
” is displayed.
Press
[Mode]
to scroll through the options for auto-zero settings (0.5d, 1d, 2d, 4d).
The normal setting is 1 scale division. The display will show “
”.
Press the
[Zero]
key to store the changed setting.
Press
[Mode]
to go back to the first technical Parameters.
Or, press
[Print]
to return to normal weighing.