- 19 -
Range
RMS Value
AC Voltage
3.000V
0.000V ~ 4.999V
30.00V
00.00V ~ 49.99V
300.0V
000.0V ~ 499.9V
DC Voltage
5.000V
0.000V ~ 9.999V
50.00V
00.00V ~ 99.99V
500.0V
000.0V ~ 999.9V
AC·DC Current
9.000mA
0.000mA~9.999mA
90.00mA
00.00mA~99.99mA
900.0mA
000.0mA~999.9mA
9.000A
0.000A~9.999A
15.00A
00.00A~15.00A
6. Range Setup
This instrument provides both auto and manual range measurement in current channel.
Refer to the following current range configuration table.
※ The default range after power on is auto range.
※ Manual range measurement is necessary when you need quick measurement response or pulse input.
When the range is set to auto, the range may be adjusted frequently if a waveform such as a pulse is
input. In such a case, set the range manually.
How to set manual range
1. Press “VOLT” and the voltage range will be modified.
2. In the above state, each time you press “RANGE,” the voltage range will be modified.
Initial (Default) Condition Auto(“Auto”sign appears)0.000V -> Manual(“Auto”sign disappears)0.000V ->
00.00V -> 000.0V -> Repeat
3. When “CURRENT” is pressed, the electric current range will be changed.
4. In the above state, each time you press “RANGE,” the electric current range will be changed.
Initial (Default) Condition Auto(“Auto”sign appears)00.00mA ->Manual(“Auto”sign disappears)00.00mA
-> 000.0mA -> 0.000A ->00.00A -> CT -> Repeat