If Balance is connected externally to PC or Data Printer via Rs232 Cable, successful or unsuccessful calibration report will be automatically generated
after the completion of Calibration process.
Calibration Report
24
Unsuccessful Calibration
When GLP OFF
When GLP ON
------------------------------
14-Jul-10
03:46PM
Aczet
Model
CY 224
Ser.no.
9930508
Ver.no.
R0.1.04
ID
1234567
------------------------------
Calibration:
External
W-ID ..................
Temperature
32.898'C
Set
+ 20.00g
External Cal Failed
------------------------------
14-Jul-09
03:46PM
Name: ................
------------------------------
Calibration
:
External
W-ID
............
Temperature
32.905'C
Set
+ 20.00g
External Cal Failed
Successful Calibration
When GLP OFF
Calibration
:
External
W-ID
............
Temperature
32.905'C
Set
+ 20.00g
Diff.
-0.01g
External Cal Done
Diff.
0.00g
When GLP ON
------------------------------
14-Jul-10
03:46PM
Aczet
Model
CY 224
Ser.no.
9930508
Ver.no.
R0.1.04
ID
1234567
------------------------------
Calibration:
External
W-ID ..................
Temperature
32.898'C
Set
+ 20.00g
Diff.
0.00g
External Cal Done
Diff.
0.00g
------------------------------
14-Jul-09
03:46PM
Name: ................
------------------------------
To obtain weighing results, the balance must be matched to the acceleration due to
gravity at its location.
Calibration is necessary
ð
Before the balance is used for the first time.
ð
At regular intervals during weighing operation.
ð
After a change in location.
Procedure
To obtain accurate results, the balance must be connected to the power supply and
allowed to warm up to the operating temperature as described on Page No
Ensure that the weighing pan is unloaded and close the doors of the draft shield (if
used). Balance should be Zero before calibration.
ð
Have required calibration weight ready
ð
Press and hold <CAL> key, display, shows “CAL Et”
ð
Release <CAL> key now.
ð
The required calibration weight value will be displayed.
ð
Place calibration weight in center of pan.
The calibration (adjustment) is finished when “CAL donE” message is displayed.
The balance is again in the weighing mode and ready for operation.
Note
: With certified balances, the calibration can be disabled after installation if
required by the national certification regulations.
The adjustment can be terminated at any time using the <CANCEL> key.
The following message appears : ‘Abort’
14
23
Calibration
4.1 External Calibration (adjusting)