4.1.6.
Laser Trigger (Green)
The Green wire is the Trigger input. It is normally left unconnected to enable the laser.
The trigger input is intended for triggering a distance measurement with an external signal
that is applied as a voltage on the green wire. A voltage between 3.0 V and 30 V is a HIGH
input and a voltage between 0 V and 1.5 V is a LOW input. It is for the user to specify a
desired delay time and a pulse flank to be selected for synchronization. The user must select
the DF measurement mode setting in the AR3000. An <Esc> character must be sent to exit
the trigger mode.
Figure 14 Wiring diagram for Trigger Input
If a switch is used, the input may need a debouncing circuit to prevent multiple triggers or
triggers on the wrong edge.
Figure 15 Debounce circuit for Trigger Input switch