![Acuity AccuRange AR1000 User Manual Download Page 16](http://html1.mh-extra.com/html/acuity/accurange-ar1000/accurange-ar1000_user-manual_2844426016.webp)
AR1000 User’s Manual
Rev 1.7 1/10
10
AR1000 accepts the command and sets the baud rate to 9600. The possible baud rates are
2400, 4800, 9600, 19200, and 38400.
3.5.2.
Sensor Output Check
If the sensor output value is in error, check that the sensor and target are stationary and stable,
that the target is at least 4 inches (0.1 m) from the sensor’s lens, and that the laser beam is
hitting the target.
The distance offset setting may alter the values output by the sensor. Reset the sensor to the
factory default to remove its effect.
The sensor may need to warm up for 5-10 minutes before reaching full accuracy. Leave it on
for a few minutes and re-check the sensor accuracy.