V. STANDARD CALIBRATION PROCEDURE

 

 

The standard CALIBRATION procedure should be performed at a minimum, every 4-hours of use, to maintain 
instrument accuracy. Calibration can be performed more frequently if desired. Additionally, once the 
DATE/TIME is set, the computer keeps track of calibration intervals and will also prompt the user to recalibrate 
every 24-hour cycle when the time reaches 00:00:01.     
 
The E20CP-III also automatically calibrates each time the sample drawer is opened.  It reads the Yellow Life 
Disk under the drawer and adjusts to a reference established during the standard calibration procedure.  

 

The E20CP-III can be configured to read samples in the Commercial Scale (the standard scale for E20CP 
models), the Gourmet Scale (the standard scale for Agtron M-BASIC models), or a user defined Custom Scale.  
 

 

The Commercial Scale offers greater dynamic range (the capacity to measure a wider range of roast 
classifications and grinds including soluble coffee powders)  

 

The Gourmet Scale offers higher resolution (shows a greater difference in scores between two roast 
classifications) 

 

The Custom Scale allows the user to arithmetically modify the Commercial Scale 

  

The E20CP-III can also be calibrated with REFERENCE COFFEE SAMPLES of known or desired score 

(see section VIII. “Non-Standard Scale Calibration”) 

 
1) Remove the two-sided Red-Black Calibration Disk set and dish from its storage box 

 
IMPORTANT: The two-sided calibration Disk & Dish set supplied with the analyzer is your permanent 
calibration reference. Disk & Dish sets are matched by serial number to a specific E20CP-III and can 
only be used with that unit.

  

 
NOTE:

 There are two adhesive labels affixed to the side of the Calibration Reference Disk dish. One label lists 

the red-side and black-side calibration reference scores; these disk values are pre-programmed into your 
analyzer at the factory. The other label shows the unit serial number and disk certification date. 

 

 

NOTE:

 Avoid touching the surface of the disks. Periodically clean the disk surfaces with a dilute solution of 

alcohol and pure water and a soft cotton cloth using very light pressure. Keep the disk set stored in the box 
provided, away from direct sunlight or heat

 

.  
● 

Do not

 leave the calibration disk set in the analyzer  

 

● 

Do not

 use the dish supplied with the calibration disk set for coffee samples  

 
IMPORTANT:

 

Always place the calibration disk into the shallow side of the dish 

 
2) Momentarily depress the [

CAL]

 key on the function keypad 

 
The display will show:                            - 

OPEN SAMPLE DRAWER 

-

 

  < SET REFERENCE MODE >

 

 
3) Fully open the sample drawer 
The display will show:                               -- 

HIGH   REFERENCE 

--

 

RED-DISK / SAMPLE 

Summary of Contents for E20CP-III

Page 1: ...E20CP III OPERATIONS MANUAL Software Version IR128 S5 0 IR128 S5 5 USB ADT 2020 Special Application Spectrophotometer AGTRON INCORPORATED 9395 Double R Boulevard Reno NV 89511 PHONE 1 775 850 4600 FAX...

Page 2: ...in a linear manner relative to the development of the collective organic chemistry The Agtron model E20CP III evaluates changes to these selected compounds by measuring their photometric response to...

Page 3: ...THE ANALYTICAL SCALE Page 12 VIII ACCESSING EDITING THE CUSTOM SCALE VARIABLES Page 13 IX IMPROVING ANALYTICAL ACCURACY AND REPEATABILITY Page 16 X NON STANDARD SCALE CALIBRATION Page 17 XI CONNECTIN...

Page 4: ...y pulling it straight out until it hits the bumper stop Look into the chassis under the drawer There are four posts sticking up from a raised platform that is affixed to the inside bottom of the chass...

Page 5: ...a 15 watt light bulb Turn the power switch to the ON position The display should activate and begin timing a 45 minute warm up period The unit model type serial number and other information will appe...

Page 6: ...keypad KEYS ANALYZER FUNCTION 0 9 Enters numeric values and access combination key functions CAL Activates calibration routine using stored high low values no editing function F CAL Displays stored h...

Page 7: ...you wish to change the Date Time displayed Depress the key The display will show YEAR YY Using the numeric keypad enter the last two digits of the year i e 2018 would be 18 The display will show MONTH...

Page 8: ...ill show AGTRON UNIT NN Use the numeric keypad to enter the two digit number to be assigned to the unit 01 99 If you wish to exit without entering an ID number key in 00 As soon as a two digit ID numb...

Page 9: ...ed with REFERENCE COFFEE SAMPLES of known or desired score see section VIII Non Standard Scale Calibration 1 Remove the two sided Red Black Calibration Disk set and dish from its storage box IMPORTANT...

Page 10: ...he sample drawer and will store a reference value The reference value stored is used automatically to recalibrate the analyzer every time the drawer is opened The display will change to LOW REFERENCE...

Page 11: ...container and thoroughly mixing the grounds to produce homogeneity It is also important that the particle specification is consistent Overfill the shallow side of the sample dish with the homogenized...

Page 12: ...S B x S C See Section VII for programming instructions To Switch Select Scales 1 Depress SCALE Each time the SCALE button is depressed selections will toggle in a loop between Agtron Scale Gourmet Sca...

Page 13: ...score offset constant VARIABLE DEFAULT VALUES LIMITS Variable A Default 0 0000 Limits 0 00000 to 0 59998 Variable B Default 1 000 Limits 0 000 to 9 876 Variable C Default 00 00 Limits 00 00 to 98 76...

Page 14: ...played depress the FNC key To edit the B variable depress the ENTER key The display will show ENTER TO SAVE EXIT B x S B 00 000 Key in the desired B value and depress the ENTER key to save and move to...

Page 15: ...ER key to save the C value exit the custom scale variable editing subroutine and return to normal analyzer function The display will show COFFEE MODE NOTE When all of the Custom Scale variables are se...

Page 16: ...f the surface compacts compresses and cakes the grind is too fine Remember what looks like an identical surface to the eye may appear quite different to the analyzer Time from Roasting Cooling Grindin...

Page 17: ...the 3 key then depress the F CAL key The display will show OPEN SAMPLE DRAWER SET REFERENCE MODE 3 Fully open the sample drawer The display will show HIGH REFERENCE RED DISK OR SAMPLE 4 Place the ligh...

Page 18: ...ibration to that scale a With the analyzer in the COFFEE MODE fully open the sample drawer and allow the analyzer to calibrate b Place the Calibration Reference disk set into the shallow side of the d...

Page 19: ...a Terminal WARNING The E20CP III does not have serial port isolation Connecting the analyzer to a non isolated serial port can permanently damage the internal micro processor Connect the analyzer seri...

Page 20: ...smitted with the notation that the auto data transmit feature is ON With auto transmit ON each time a reading is taken the product score with time and date will be transmitted b Depress the 6 SEND DAT...

Page 21: ...E USING THE NUMERIC KEYPAD AND DEPRESS THE ENTER KEY TO ENTER THAT VALUE 8 FULLY OPEN THE SAMPLE DRAWER CAREFULLY PLACE THE DARKER SAMPLE INTO THE SAMPLE POSITION AND FULLY CLOSE THE DRAWER 9 KEY IN T...

Page 22: ...e disk set is not in use Never leave the disk set in the analyzer CLEANING THE ANALYZER INTERIOR Unplug the analyzer Remove the LIFE DISK from under the sample drawer Clean the life disk with alcohol...

Page 23: ...56 9 75 70 53 1 NO TILE Medium Light 65 49 4 65 60 45 7 NO TILE Medium 55 42 0 55 50 38 3 NO TILE Medium Dark 45 34 6 45 40 30 8 NO TILE Dark 35 27 1 35 30 23 4 NO TILE Very Dark 25 19 7 25 Extremely...

Page 24: ...CTIONS Contact Agtron for a Return for Repair authorization number RFR before sending equipment to Agtron for service Analyzers returned without an RFR will not be received Repack the equipment in the...

Page 25: ...25 XV CE CERTIFICATE DECLARATION of CONFORMITY...

Reviews: