Actegy REVITIVE 2836AA User Manual Download Page 12

22

23

h

  Press the power button on the device 

or

 on the    

  remote control to turn REVITIVE on.    

 

 

  You will hear a beep and the display lights will light up.   
  The      time will read 20 minutes and       Intensity will  
  read 00.         
  To turn off REVITIVE at any time during operation,   

  press the power button on the device 

or

 the remote  

 control.

i

  Set the time for the number of minutes you want to  

  use REVITIVE. The default time for the Arthritis Pads  

 

is 

20 

minutes. 

 

      

  You can use the buttons on the device 

or

    

 

 

remote 

control. 

      

  Press the     (+) Time Contol to increase the time.   
  Press the     (-) Time Contol to decrease the time.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

NB 

The Arthritis Pads are only suitable for use with the REVITIVE Arthritis-

Knee device and will not be effective for any other areas of the body. 

Do NOT use the Arthritis Pads on your head, face, neck or chest.  

j

  Press the       (+) Intensity Control once to start the   

 stimulation.
  Continue to increase the intensity until you can feel or  

  see the muscles in your thigh contract and relax. You  

  can set the intensity between 1-99.

  You can use the buttons on the device 

or

 remote    

 

control. 

       

  Press the (+) button to increase the level. 

 

 

  Press the (–) button to decrease the level.

  Use an intensity that provides you with strong but    

  comfortable muscle contractions and note that the   

  intensity required may vary from day to day - eg    

  using it on 25 one day, does not necessarily mean    

  you will need to use it at 25 the next day. The required  

  intensity could be higher or lower than previous usage. 

  It is normal to feel tingling or varying sensations in   

  your thigh muscles. 

Instructions for use

Adjusting the intensity level 

 -  At the start of therapy there will be  

30 seconds of stimulation to help you set the right intensity level. This is 

followed by a rest period of 30 seconds, when you will feel no stimulation.

For the remainder of the therapy session you will feel 10 seconds of 

stimulation, followed by a 50 second rest period which repeats until the 

end of therapy.

Adjust the intensity when you can feel the stimulation being applied to 

your thigh. 

It is not possible to adjust the intensity during the rest period.

2910_IFU02_12517049.indd   23

22/07/2016   09:49

Summary of Contents for REVITIVE 2836AA

Page 1: ...r 2836AA User s Manual PLEASE READ THE USER S MANUAL CAREFULLY BEFORE USING THIS PRODUCT USE ONLY AS DIRECTED If symptoms persist consult your healthcare professional 2910_IFU02_12517049 indd 1 22 07...

Page 2: ...tructions for use 13 27 How long should I use REVITIVE 13 Step 1 Setting up REVITIVE for the first time 13 Step 2 Setting up REVITIVE for use 14 Step 3 Powering REVITIVE 15 16 Step 4 Using the foot pa...

Page 3: ...F G J I H Carry handle Control panel display Power ON OFF Foot pads AC DC power adaptor REVITIVE C D Arthritis Electrode Pads x 2 pairs Arthritis Pad cord Remote control AAA batteries x 2 Controls of...

Page 4: ...while you are seated The EMS causes muscles in your calf and foot to contract which helps to increase circulation in your legs REVITIVE has been developed so that it is comfortable for you to use It f...

Page 5: ...ritis Pads with others Do not use Arthritis Electrode Pads in conjunction with a brace or cast without first consulting with your doctor Do not use Arthritis Electrode Pads after their use by date IMP...

Page 6: ...recautions Do not stand on the machine Use only when sitting down Use the device only for its intended purpose Do not expose the device to extreme heat Do not spill liquid on the device or its accesso...

Page 7: ...pages 8 12 If in doubt consult your doctor before using the product How long should I use REVITIVE It is recommended to use REVITIVE foot pads for 20 30 minutes each day seven days a week It is recomm...

Page 8: ...the IsoRocker Disabler and press into place on the bottom of the device ie flat Full Battery Low Battery Instructions for use STEP 3 POWERING REVITIVE REVITIVE can be powered from either the rechargea...

Page 9: ...th your knees at a 90 degree angle right angle b Place REVITIVE on the floor in front of you REVITIVE is designed to be used while seated Never stand on the REVITIVE unit c Place both bare feet on the...

Page 10: ...ess the button to increase the level Press the button to decrease the level Instructions for use h Use an intensity that provides you with strong but comfortable muscle contractions and note that the...

Page 11: ...is recommended that the Arthritis Pads be used on the affected leg knee for 20 minutes once a day 5 days per week Find a routine that works for you perhaps using the Arthritis Pads Monday Friday and...

Page 12: ...ax You can set the intensity between 1 99 You can use the buttons on the device or remote control Press the button to increase the level Press the button to decrease the level Use an intensity that pr...

Page 13: ...u feel pain or discomfort lower the intensity of the Arthritis Pads using either the Intensity Control on the control panel of the device or on the remote control You can also choose to stop therapy b...

Page 14: ...device while it is charging but the battery indicator will not show Low Battery Full Battery STEP 7 CLEANING AND STORING REVITIVE a Ensure REVITIVE is switched off b Wipe down the foot pads with a so...

Page 15: ...l cannot feel the stimulation then please contact your authorised dealer Troubleshooting Problem Possible Cause Solution No power or lights to REVITIVE when it is switched on AC Adaptor not switched o...

Page 16: ...mation including Frequently Asked Questions please visit www revitive com Name of product REVITIVE Arthritis Knee Model 2836AA Frequency 1Hz 50Hz Output current Max 15mA Weight 1 5kg Dimensions with I...

Page 17: ...exclusions set out below see Exclusions the faulty device will then be replaced and dispatched usually within 14 working days of receipt If for any reason this item is replaced during the 2 year guara...

Page 18: ...34 35 Notes 2910_IFU02_12517049 indd 34 35 22 07 2016 09 49...

Page 19: ...e registered trademarks of Actegy Ltd The REVITIVE product is a proprietary design and is protected by applicable design laws AW 2910_IFU02_12517049 P D 08 2016 Actegy Ltd REFLEX Cain Road Bracknell R...

Reviews: