INSTRUCTION MANUAL
4
•
Disconnect the simulator power after use
•
CAUTION! After disconnecting the simulator power, allow 10 to 15 minutes for
the heating element to cool before detaching the top housing
•
Empty out the solution container at the end of the work day
•
Store the simulator in an environment of 5 to 40 °C and 10 to 85% RH
ALCOHOL REFERENCE SOLUTION
•
Use the solution concentration specified in the manual for breath testing
devices
•
Replace solution every 5 days or 20 tests
•
Use of solution over time affects alcohol concentration
•
Do not use a solution bottle with a broken seal, or an expired bottle
•
Never use artificial methods to reheat or cool solution, or the solution
container
•
Keep solution at room temperature
•
Do not freeze or refrigerate solution
•
Do not ingest solution
•
If solution is ingested, do not induce vomiting; contact your local poison
control centre
•
Keep solution away from eyes
•
If solution comes in contact with eyes, flush eyes with water; if irritation
continues, contact your local poison control centre
•
Refer to your local environmental regulations for more information about safe
solution-disposal amounts
•
Dispose solution down a drain
CLEANING THE ALCOSIM BREATH ALCOHOL
SIMULATOR
•
Disconnect the ALCOSIM breath alcohol simulator power before cleaning
•
CAUTION! After disconnecting the ALCOSIM simulator power, allow 10 to 15
minutes for the heating element to cool off before detaching the top housing
•
Do not submerge the top housing in water
•
Clean the top housing by wiping with a water dampened cloth
•
Wash the solution container with plain water and dry with paper towel