ACS ALCOSIM Instruction Manual Download Page 10

ALCOSIM

7

TROUBLESHOOTING

The following table lists error codes that may appear on the screen of the 

ALCOSIM simulator. The error codes are accompanied by a periodic beep.

ERROR CODE DESCRIPTION

WHAT TO DO

Er 11

Motor malfunction.

Switch off the ALCOSIM simulator, 

wait 5 seconds and switch it on 

again.
If the error message persists, return 

the simulator to an authorized 

service provider.

Er 13

Transistor malfunction.

Er 21

Temperature sensor 1 has 

no current.

Er 22

Temperature sensor 1 short 

circuit.

Er 23

Temperature sensor 1 leak.

Er 25

Electric potential meter 1 

damage.

Er 31

Temperature sensor 2 has 

no current.

Er 32

Temperature sensor 2 short 

circuit.

Er 33

Temperature sensor 2 leak.

Er 35

Electric potential meter 2 

damage.

Er 41

Heater malfunction.

Er 52

The water level in the jar is 

lower than normal. 
The power supply used is 

not adequate.

Fill the jar to the correct liquid volume 

of 0.5L.
Use the power supply provided with the 

ALCOSIM.

Er 54

Alcohol reference solution 

temperature above 34 °C 

for longer than 

3 minutes.

The ALCOSIM simulator 

attempts to restore 

temperature to 34 °C. If it 

cannot, the heater disables 

and an 

Er 54

 message 

appears.

NOTE: Use the simulator within room 

temperature environments (23 °C ± 

5 °C).

 

Turn the simulator off in order for the 

alcohol reference solution to cool down. 

Turn it on again when appropriate.

Return the simulator to an authorized 

service provider if the error messages 

persist and ambient temperature is not 

the cause.

 

Er 55

Alcohol reference solution 

temperature above 37 °C.
The heater automatically 

disables and an 

Er 55

 

message appears.

Summary of Contents for ALCOSIM

Page 1: ...Instruction manual...

Page 2: ...tional Inc and are used under license Alcohol Countermeasure Systems is trading style of Alcohol Countermeasure Systems International Inc 2012 2014 Alcohol Countermeasure Systems The information discl...

Page 3: ...sistance 2 ALCOSIM breath alcohol simulator Overview 2 Safety and precautions 3 ALCOSIM setup usage and disassembly 3 Alcohol reference solution 4 Cleaning the ALCOSIM breath alcohol simulator 4 Calib...

Page 4: ...r Maintains a precise temperature at 34 C 0 02 C Brushless DC motor operating at 24VDC capable of running 24 7 System for detecting motor failures due to mechanical and electrical defect Audio and vis...

Page 5: ...te and verify breath alcohol testing devices The simulator is a wet bath calibration verification device FRONT Illuminated ON OFF switch Air outlet tube Baf e plate 500 mL solution fill line Wide view...

Page 6: ...urface avoid contact with the heating element Place the ALCOSIM simulator on a flat surface free from obstruction Do not expose the simulator to direct sunlight for extended periods of time Do not use...

Page 7: ...n at room temperature Do not freeze or refrigerate solution Do not ingest solution If solution is ingested do not induce vomiting contact your local poison control centre Keep solution away from eyes...

Page 8: ...ACS 79 007600 AC power cord part number depends on your region Power supply 24VDC 2 5A ACS 07 000075 50cm of vinyl tubing ACS 70 000002 Liquid Trap Mouthpiece ACS 79 001955 Round mouthpiece ACS 95 00...

Page 9: ...r in the solution In case of bubbling check that the ALCOSIM head is tightened properly Return the ALCOSIM to an authorized Service Provider if the problem persists 6 Connect the power cable to the si...

Page 10: ...35 Electric potential meter 2 damage Er 41 Heater malfunction Er 52 The water level in the jar is lower than normal The power supply used is not adequate Fill the jar to the correct liquid volume of...

Page 11: ...UAL 8 INDEX A Alcohol Reference Solution 4 5 ALCOSIM simulator Parts 2 Product features 1 Setup and operations 6 Assistance 2 C Calibration materials 5 E Error codes 7 S Safety and precautions 3 T Tro...

Page 12: ...IRM ENG 60 000180 I 2012 2014...

Reviews: