- find the maximal signal from the second reference reflector and adjust it in the same manner, at that the first anchor point
shall be out of the strobe field (Figure 27);
- press the
key, at that the second anchor point will appear on the peak value of the signal;To delete the anchor point,
press the F6
key.
Note:
several anchor points will be deleted according to their creation sequence.
- create the anchor points for all reference reflectors;
- press the
key;
- confirm or cancel the DAC curve (Figure 28).
Press the
key to exit the configuration window not saving the settings.