Ultrasonic flaw detector А1212 MASTER
Operation Manual
21
ACOUSTIC
CONTROL
SYSTEMS
Figure 14
2
2.3.1.2 Parameters of the signal
The screen when setting the signal parame-
ters is presented in the Figure 14.
The names of the signal parameters and their
permissible values are presented in the Table 7.
Table 7
Parameter
Value
Description
Gain, dB
from 0 to 100
Select the gain in the receive path
Gain step, dB
1 / 6 / 10
Select the gain step
Impulse, V
20 / 100 / 200
Select amplitude of the sounding impulse
Excitation impulse, period
from 0.5 to 5.0 with step 0.5 Model the sounding signal
Frequency of the snapshots, Hz 25 / 50 / 250 / 500 / 1000
Select the regularity of information visualisation on the screen