Level Gauge A1207D
User Manual
18
ACOUSTIC
CONTROL
SYSTEMS
8
TROUBLE-
SHOOTING
Problem
Possible Reason(s)
Action
Incorrect measurement
during functionality check
Calibration block has not adequately
been used.
- Disconnect the probe from the calibration block.
- Verify the quantity of couplant
- The coupling surface must be clean, not damaged
and without adherent foreign particles.
- Establish the connection again.
- Read the correction value for temperature from the
table, and check the result.
Incorrect measurement
Improper couplant.
Replace couplant
Incorrect measurement
Excessive quantity of couplant.
Remove some couplant.
No signal
Insufficient quantity of couplant.
Add some couplant.
Incorrect measurement,
no result
The probe has been introduced in a ske-
wed manner into the guide of the oil pan.
- Detach the probe
- Insert it again, correctly oriented.
No signal
Damaged probe
Probe must be replaced.
No signal
Damaged probe cable
Cable must be replaced.
Damaged casing, re-
chargeable battery, display
and electronics
Device has been dropped
Send the device back for repair in its original box,
or hand it over to a designated collection point for
recycling.
The product must not be
disposed of with the domestic waste.
Summary of Contents for A1207D
Page 20: ...Level Gauge A1207D User Manual 20 ACOUSTIC CONTROL SYSTEMS NOTES ...
Page 21: ...Level Gauge A1207D User Manual 21 ACOUSTIC CONTROL SYSTEMS NOTES ...
Page 22: ...Level Gauge A1207D User Manual 22 ACOUSTIC CONTROL SYSTEMS NOTES ...
Page 23: ...Level Gauge A1207D User Manual 23 ACOUSTIC CONTROL SYSTEMS ...