2.2 Real-time averaging mode (AVG option)
24
Acqiris SA240P User's Manual
It is possible to use the NSA on negative polarity pulse. By enabling the channel data inversion
capability, the signal will be inverted before applying the NSA settings. See
.
The NSA applies on the signal after the baseline stabilization. The baseline stabilization
parameters must be configured before enabling the NSA.
Parameters
Enabled
: Specifies whether the noise suppressed accumulation is active. This attribute affects card
behavior only when the acquisition mode attribute is set to averager.
Threshold
: Specifies the threshold of the noise suppressed accumulation. Each data value must
exceed the threshold value to be entered into the sum. This parameter is defined as signed left-
aligned 16-bit ADC code, and in the range of [-32768, +32767]. Data values below the Threshold are
set to -32768 (or to Digital Offset if the Baseline Stabilization is enabled).
NoiseBase
: Specifies the noise base value for the noise suppressed accumulation. The noise base
is subtracted from data values which are higher than the configured threshold. This parameter is
defined as signed left-aligned 16-bit ADC code, and in the range of:
[Threshold - 32767, Threshold] if Threshold ≥ 0
[Threshold, Thr32768] if Threshold < 0
Sequence
When enabling the NSA, the recommended sequence is the following:
1. Configuring the channel and acquisition parameters (voltage range, offset, sampling rate,
record size, ...)
To configure the settings the number of average and the number of record should be
set to 1.
2. Inverting data (optional): The data inversion is required for negative pulses
3. Enabling the baseline stabilization: The baseline stabilization is applied on the signal after the
data inversion.
Mode
: Continuous
Pulse Polarity
: Positive (only positive is supported)
Pulse Threshold
: set the Pulse Threshold just above the signal baseline
Digital Offset
: apply a digital offset on the sample data
4. Enabling the NSA: The NSA is applied after the baseline stabilization.
NSA
: enabled
Threshold
: all samples below the Thresold are set to
Digital Offset
(or -32768 with
no baseline stabilization)
Noise Base
: all samples above the
Thresold
are set to
Threshold - NoiseBase
Please contact [email protected] for guidelines to setup NSA.