Serinus 60 User Manual 3.0
Page 56
3.4.26
Digital Input Menu
Main Menu
Communications Menu
Digital Input Menu
This menu is used to remotely trigger zero and span calibrations. This is done by assigning the eight
digital inputs with one of the following commands.
DI N (Pin X)
Associates an action with a digital input. There are eight
available digital inputs (the pin numbers are for the 25-pin
connector). Each one can have one of the following associated
actions, triggered when the corresponding digital input goes
to the
Active
state:
Disabled:
No action (this digital input does nothing) the pin
is forced to Active Low.
Do Span:
Used to perform a span check When activated the
instrument sets the
Cal. Mode
to
Span (
refer to Section
Do Zero:
Used to perform a zero When activated the
instrument sets the
Cal. Mode
to
Zero
(refer to Section
Active
[DO Span, Do Zero]
Each pin can be set to be active
High
or
Low
. Active High
means that the event will be triggered when the line is pulled
to 5V. Active Low means that the event will be triggered when
the line is pulled to 0V.
Example
Here is a typical configuration between an instrument and either a data logger or calibrator (master
device):
1.
Set the jumper JP1 to 5V position (refer to Section 4.4.3).
2.
Connect one of the master devices digital output signals to pin 18 and the ground signal to pin 5
of the instrument’s analog/digital 25 pin female connector (refer to Figure 18).
3.
Program master device to output 0 volts to pin 18 when a span is desired.
4.
In the instrument’s
Digital Input Menu
assign
DI 0
Do Span
.
5.
Set the
Active
state to
Low
6.
The same procedure can be followed to also activate zero calibrations. Pin 6 of the instrument’s
analog/digital 25 pin female connector can be connected to one of the other master devices
digital outputs and the instrument can be set so
DI 1
is assigned to
Do Zero
.
3.4.27
Digital Output Menu
Main Menu
Communications Menu
Digital Output Menu
This allows the instrument to trigger external alarms in response to certain events.
Summary of Contents for Met One Instruments GAS-1060
Page 10: ...Serinus 60 User Manual 3 0 Page 10 This page is intentionally blank...
Page 18: ...Serinus 60 User Manual 3 0 Page 18 This page is intentionally blank...
Page 35: ...Installation Page 35 Figure 7 Switching the Battery On Off...
Page 36: ...Serinus 60 User Manual 3 0 Page 36 This page is intentionally blank...
Page 102: ...Serinus 60 User Manual 3 0 Page 102 This page is intentionally blank...
Page 134: ...Serinus 60 User Manual 3 0 Page 134 9 5 Plumbing Schematic PN D020060...
Page 135: ...Parts List Page 135 9 6 Block Wiring Schematic PN D020111...
Page 136: ...Serinus 60 User Manual 3 0 Page 136 9 7 Measurement Cell Side Section PN H015000...
Page 137: ...Parts List Page 137...
Page 139: ...Parts List Page 139 9 9 Valve Assembly Exploded View PN H010042...
Page 140: ...Serinus 60 User Manual 3 0 Page 140 9 10 High Pressure Valve Exploded View PN H050043...