![ACOEM Met One Instruments GAS-1060 User Manual Download Page 68](http://html1.mh-extra.com/html/acoem/met-one-instruments-gas-1060/met-one-instruments-gas-1060_user-manual_459535068.webp)
Serinus 60 User Manual 3.0
Page 68
4.4.2
Analog Inputs
The instrument is also equipped with three analog inputs with resolution of 15 bits plus polarity,
accepting a voltage between 0-5V. These go directly to the microprocessor and should be protected
to ensure static/high voltage does damage the main controller PCA (instrument warranty does not
cover damage from external inputs).
4.4.3
Digital Status Inputs
The instrument is equipped with eight logic level inputs for the external control of the instrument
such as Zero or Span sequences. Each input has a terminating resistor which can be either PULL UP or
PULL DOWN. This is set using the jumper JP1 on the back panel PCA (refer to Figure 17).
4.4.4
Digital Status Outputs
The instrument is equipped with eight open collector outputs which will convey instrument status
condition warning alarms such as no flow, sample mode, etc. Two of the digital outputs can be set so
that there is +5V and +12V available on the 25 pin connector for control purposes, instead of digital
outputs 0 and 1.
In the default jumper locations (refer to Figure 10) these two outputs will function normally as open
collector outputs. If moved to the position closer to the 25 pin connector then the DO 0 will supply
+12V and DO 1 will 5V.
The +12V and +5V supplies are limited to about 100mA each.
Each digital output is limited to a maximum of 400mA. The total combined currents should not
exceed 2A.
Figure 17 – 25 Pin Rear Panel PCA (Default Jumpers Highlighted)
Summary of Contents for Met One Instruments GAS-1060
Page 10: ...Serinus 60 User Manual 3 0 Page 10 This page is intentionally blank...
Page 18: ...Serinus 60 User Manual 3 0 Page 18 This page is intentionally blank...
Page 35: ...Installation Page 35 Figure 7 Switching the Battery On Off...
Page 36: ...Serinus 60 User Manual 3 0 Page 36 This page is intentionally blank...
Page 102: ...Serinus 60 User Manual 3 0 Page 102 This page is intentionally blank...
Page 134: ...Serinus 60 User Manual 3 0 Page 134 9 5 Plumbing Schematic PN D020060...
Page 135: ...Parts List Page 135 9 6 Block Wiring Schematic PN D020111...
Page 136: ...Serinus 60 User Manual 3 0 Page 136 9 7 Measurement Cell Side Section PN H015000...
Page 137: ...Parts List Page 137...
Page 139: ...Parts List Page 139 9 9 Valve Assembly Exploded View PN H010042...
Page 140: ...Serinus 60 User Manual 3 0 Page 140 9 10 High Pressure Valve Exploded View PN H050043...