Figure 18 – External 25 pin I/O Individual Pin Descriptions
CAUTION
The analog and digital inputs and outputs are rated to CAT I.
Exceeding 12VDC or drawing greater than 400mA on a single output or a total
greater than 2A across the eight outputs can permanently damage the instrument
and void the warranty.
4.5
Logging Data
When the user receives the instrument from the factory it will have a default set of parameters
already setup in the internal data logger. These select few parameters have been chosen for their
relevance in assisting in troubleshooting the instrument.
4.5.1
Configure Instrument Internal Logging
In order to log data you must first specify a data logging interval. This is how often data will be logged
to the USB memory stick. You can log a maximum of 12 parameters. These parameters can be
selected by name or by parameter number using the parameter list as a reference (refer to Table 19).
Procedure
1.
Open -
Main Menu
Communications Menu
Data Logging Menu
(refer to Section 3.4.22).
2.
Select -
Data Log Interval
(adjust to the desired value).
Summary of Contents for Met One Instruments GAS-1060
Page 10: ...Serinus 60 User Manual 3 0 Page 10 This page is intentionally blank...
Page 18: ...Serinus 60 User Manual 3 0 Page 18 This page is intentionally blank...
Page 35: ...Installation Page 35 Figure 7 Switching the Battery On Off...
Page 36: ...Serinus 60 User Manual 3 0 Page 36 This page is intentionally blank...
Page 102: ...Serinus 60 User Manual 3 0 Page 102 This page is intentionally blank...
Page 134: ...Serinus 60 User Manual 3 0 Page 134 9 5 Plumbing Schematic PN D020060...
Page 135: ...Parts List Page 135 9 6 Block Wiring Schematic PN D020111...
Page 136: ...Serinus 60 User Manual 3 0 Page 136 9 7 Measurement Cell Side Section PN H015000...
Page 137: ...Parts List Page 137...
Page 139: ...Parts List Page 139 9 9 Valve Assembly Exploded View PN H010042...
Page 140: ...Serinus 60 User Manual 3 0 Page 140 9 10 High Pressure Valve Exploded View PN H050043...