![ACKERMANN 16-2050 User Manual Download Page 18](http://html.mh-extra.com/html/ackermann/16-2050/16-2050_user-manual_458095018.webp)
SHAVER SYSTEM
Ackermann
Page: 18 // 28
6.4. REUSABIlITY
The individual components can be reused as long as they perform adequately and the instruments can be fixed properly,
provided that they are used with due care and reprocessed, inspected and packed correctly. Any (further) use of damaged
or contaminated instruments is on the sole responsibility of the user. The durability of the product largely depends
on careful handling during use and reprocessing, which is why the number of possible reprocessing cycles cannot be
determined. The manufacturer declines all liability in case of non-observance of these instructions.
7. INSPECTIONS, SERvICE, TEChNICAl SUPPORT
7.1. REPlACING fUSES
WARNING!
follow these steps before replacing the fuses:
• turn the device off using the power switch “POWER ON/Off” (15) located on the rear panel,
• disconnect the power cord from the power socket and then from the socket of the device (15). When equipment was
switched on, please wait 3-4 minutes for disconnecting.
Replacing fuses is to be performed only if they are damaged.
Use only such fuses as indicated in Table 5.
To replace the fuses:
• remove defective fuses. Fuses are located in the drawer (17) installed in the power socket (16) of the device,
• new time-delay fuses, the values of which should be selected on the basis of Table 5, should be inserted in the slots in the
fuse drawer.
Table 5. Values of fuses intended for specific models of the Control Unit of Shaver System
:
Pos.
Type of Control Unit
value and type of fuse
1
# 16-2050
2x T 1,25A L 250 V
• Slide the fuse drawer (17) into the power socket of the device (16). Proper installation will be signalled by an audible “click”
, the sound of latches closing in the drawer locking mechanism.
• Connect the power cord. The device can be restarted.
7.2. PERIODIC INSPECTIONS Of ThE DEvICE
For long-lasting and trouble-free operation of the device, the manufacturer imposes obligatory
inspections, which should take place at intervals of no less than every 12 months (once a year).
The inspection must be conducted by the manufacturer or its authorised service provider. See service provider data
in the last page.