Ackermann
Page: 11 // 28
SHAVER SYSTEM
The maximum speed for Shaver System is 8000 rpm.
Attention:
Handpieces must have come to a standstill before they can be removed from the socket (2) and/or (3) of the
Control Unit. Hold Handpieces only by their case (22) during use.
4.4. CONNECTION Of WIRED fOOTSWITCh TO CONTROl UNIT
• The plug (38) at the end of the Footswitch cable has to be connected to the “FOOTSWITCH” (4) socket at the front panel of
the Control Unit. Only one Footswitch can be connected.
• The plug (38) of the Footswitch is marked with a notch and a red dot. When connecting the plug, make sure that the red
dot on the plug is in the same position as the red dot on the socket (4) of the Control Unit. The pins inside the plug might
otherwise get damaged during connection.
• To avoid damage, make sure to pull at the plug (38) itself and not the cable when removing the plug. When the Control Unit
is activated (see chapter 4.6.), the Handpiece can be operated via the Footswitch.
4.5. CONNECTION Of WIRElESS fOOTSWITCh TO CONTROl UNIT
• The plug (38) has to be connected to the “FOOTSWITCH” (4) socket at the front panel of the Control Unit. Only one
Footswitch can be connected. Before connecting the plug (38) to the “FOOTSWITCH” (4) socket please make sure that the
three-number-codes on the receiver and the Footswitch console are the same, otherwise the Footswitch will not work.
• The plug (38) of the Footswitch is marked with a notch and a red dot. When connecting the plug, make sure that the red
dot on the plug is in the same position as the red dot on the socket (4) of the Control Unit. The pins inside the plug might
otherwise get damaged during connection.
• When the Control Unit is activated (see chapter 4.6.), the Handpiece can be operated via the Footswitch.
Please adhere to the following steps to initialize the Footswitch work:
• Turn the Footswitch on by pressing any button. The connection between the wireless Footswitch console and the receiver
(39) will be established automatically, which may take 5 sec. and in extreme cases up to 20 sec. (time of receiver (39)
verification increases e.g. because of many Bluetooth devices around). Proper data transmission is signaled by flashing,
Connection LED (42).
• Before starting the work, make sure that the accumulator is appropriately charged. If the wireless Footswitch console
is not connected with the receiver (39), the accumulator level is displayed till the connection with the receiver (39) will be
established or till the Footswitch console will be switched to stand-by mode. If the Footswitch console is connected with
the receiver (39), to check the accumulator level please press “TOOL”(33) button for about 3 seconds, the current level is
displayed on the three, green diodes (43). Diodes are turned after 10 sec. off.
• The communication range between the Footswitch console and the receiver (39) is from 0 to 3m. It is recommended
to place no obstructions on the line Footswitch console – the receiver (39), such as: metal elements connected to GND
potential, thick-walled materials absorbing radio waves frequency 2,4 GHz, that can reduce transmitted signal power.
• After breaking the communication (for example in the case of too large distance from the receiver (39)), establishing the
new connection can take 7 sec., in extreme cases up to 20 sec.
• The Footswitch is automatically switched to stand-by mode, after turning off the Control Unit with Mains switch POWER
ON/OFF (15), to save the accumulator energy. The communication is stopped. In both cases, to activate the Footswitch
(after turning off or longer not using the Footswitch) press any button on the Footswitch console. The communication will
be established again.
• To charge the accumulator please connect the battery charger (45) to the socket (41) on the Footswitch console. While
charging, connection between the Footswitch console and the receiver (39) is stopped. ( the Footswitch is inactive).
• The average charging time is 4 h and is dependent on the discharging level.
• If the red Error Diode (44) is turned on continuously then an incorrect connection of the charger is signalized or a wrong
type of charger was used.
• If the Error Diode (44) is pulsating the transmission errors related to disconnection of the receiver (39) in the Console are
signalized.
The footswitch should be recharged only outside operating and treatment rooms. The charger attached to the set is
not a medical purpose device.