USER MANUAL 16-2040N / 16-2041N / 16-2042N
Rev. 07/2016
Guide and declaration by manufacturer – electromagnetic immunity
This insufflator is designed to be used within the electromagnetic environment specified below. The user should ensure
that it is indeed used within this environment
IMMUNITY
TEST
CEI 60601
SEVERITY LEVEL
COMPLIANCE
LEVEL
ELECTROMAGNETIC ENVIRONMENT-GUIDE
Conducted RF
EN 61000-4-6
Radiated RF
EN 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3V
3V/m
Portable and mobile RF communication equipment
should be used no closer to any part of the camera
including cables, than the recommended seperation
distance calculated from the frequency of the
transmitter.
Recommended separation distance
80 MHz to 800 MHz
800 MHz to 2.5 GHz
Where
P
is the maximum output power of the
transmitter, in Watts (W), assigned by the
manufacturer of the transmitter and the recommended
separation distance (d) in meters (m).
The field levels emitted by immobile RF transmitters –
which are to be established by an electromagnetic
measurement at the site – must be lower than the
compliance level in each frequency band.
Interference can occur nearby devices bearing the
following symbol:
Note 1 At 80 MHz and 800 MHz, the higher frequency band applies.
Note 2 These recommendations may not apply in all situations. The propagation of electromagnetic waves is modified
by absorption and reflection due to structures, objects and people.
a
The field level of immobile transmitters, such as radio telephone base stations (cellular and cordless) and mobile
terrestrial radio systems, amateur radio systems, AM/FM radio communication systems, and TV systems cannot be
evaluated theoretically with precision. To ascertain the electromagnetic environment due to immobile RF transmitters, a
site measurement must be performed. If a field level measured within the environment in which this insufflator is used
exceeds the above applicable compliance levels, check that this insufflator operates in a satisfactory manner. If
abnormal operation is observed, additional measurements should be taken, such as reorientation or relocation of the
reference system.
b
Outside the frequency band of 150 kHz to 80 MHz, the field level should be less than 3 V/m
PAGE 20
OF 21