![ACKERMANN 16-2040N User Manual Download Page 18](http://html1.mh-extra.com/html/ackermann/16-2040n/16-2040n_user-manual_458090018.webp)
USER MANUAL 16-2040N / 16-2041N / 16-2042N
Rev. 07/2016
• External desufflation valve
• Option of automatic low/high flow switching
Interface:
• 5.7-inch LCD screen showing: immediate flow, intra-cavity pressure, blow-off pressure, total volume of CO2
used and low levels in the cylinder
• Thumbwheel navigation and validation system
• Internal warning data memory
Power supplies:
• Types of power supply: 100-230 V AC 50-60 Hz
• Protection by fuses: 2 x 2.5 AT – 250V time-lag fuses - UR
• Power used: 40VA
Mechanics:
• Dimensions: L=300mm, D=340mm, H=145mm
• Weight: 5 kg
Operating – transportation and Storage Environment:
• Operating temperature range: +10°C / +40°C
• Operating relative humidity range: 30 to 75 %
• Transport and storage temperature range: +10°C / +40°C
• Transport and storage relative humidity range: 20% to 85%
• Operating, transport and storage atmospheric pressure range: 700hPa to 1060hPa
Standards:
• Electrical protection: class 1 type BF
• Conforms to European directive 93/42/CEE on medical devices: class IIa
• Conforms to standard IEC 60 601–1; UL 60601-1 ; CSA 60601-1
• Not protected against water (IPX0)
• Not adapted for use in the presence of a flammable anesthetic mixture with air, oxygen or nitrous oxide
Ancillary equipment:
• Main power cable
• Open-ended spanner intended for CO2 high-pressure feed hose connector.
• 4 sterile, disposable external desufflation tubes fitted with a viral filter.
PAGE 18
OF 21