FUNCTION SETTINGS 

 

1.

 

Press [ON/OFF] to turn on the scale, the display will show flash digits and     
then --------,

 press 

[PRINT]

, the display will show 

b xxx BAU

 value (press 

[UNIT]

 to select among 1200, 2400, 4800, 9600), our default setting is 9600. 

Press 

[ZERO] 

to enter Parity mode, the display will show 

P  XXX PAR

, press 

[UNIT] to select odd,even,none (odd stands for 7 data bits with odd parity, even 

stands for 7 data bits with even parity, none stands for 8 data bit without parity)our 

default setting is none. 

Press 

[ZERO] 

to enter Print mode, the display will show n

 xxx Mod

, press [UNIT] 

to select ON, OFF, KEY, STB (OFF stands for serial data output disabled, KEY 

stands for demand mode by press 

[PRINT], STB 

stands for automatic print when 

scale is stable, ON stands for continuous print, our default setting is STB. Press 

[ZERO] again to return to weighing mode. 

 

2. Press [ON/OFF] to turn on the scale, the display will show flash digits and 
then ----------,

 press 

[UNIT]

 to enter unit selecting function, the display will show 

ON X (X stands for the units: g, oz, lb, dwt…tola...) press 

[UNIT]

 to select ON 

(activate the selected unit) or OFF (inactivate the selected unit), press [ZERO] again 

to return to weighing mode. 

 

3. Switch the power to “ON”

 and than press [ON/OFF] to turn on the scale

The 

display shows F-E-9 to 0 and than “- - - - - - -”|

press [PCS] key to enter ZERO 

track function setting

display shows 

ZRD x.xd

press [UNIT] key to select ZERO 

track value

0.0d-0.5d-1.0d-1.5d-2.0d

; press[ZERO]key to enter ZERO track 

function setting

display shows 

ZRT x.xd

press [UNIT] key to select ZERO track 

time

05. to 4.0 seconds; press [ZERO] key to enter ZERO setting

display shows 

RZD XXX

, press [UNIT] key to select ZERO value

0.8d-1.8d-2.8d-3.8d

;press 

[ZERO] key to enter 

FIL

 setting, press [UNIT] key to select filter range amend 

(0-1-2-3-4-5-6-7),”2”is default setting; press [ZERO] key to select zero range when 

turn on the scale, press [UNIT] key to select percent (

PZR

 2,    PZR 5, PZR 10. 

PZR 20); press [ZERO] key to set zero range, press [UNIT] key to select percent 

(

KZR

 3,4,5,10); press [ZERO] key to set 

TARE range

, press [UNIT] key select 

percent (

KTR

 50 or KTR 100); press [ZERO] key to enter overload protection 

setting, press [UNIT] key to select overload protection value (9d OVR-1 OVR-2 

OVR-10 OVR); press [ZERO] key to set beep function, press [UNIT] key to choose 

turn on or turn off the beep, “b2 ON” means activate the beep, “b2 OFF” means 

inactivate the beep; press [ZERO] key to unite the function of TARE and ZREO, 

press [UNIT]  to switch between them. “ZT ON” means unite function activate, 

“ZT OFF” means unite function inactivate. press [ZERO] key to enter auto off  

function setting

display shows A    xxx

press [UNIT] key to select auto off time

3,5,10,off

. Press 

[ZERO]

 to select backlight, the display will show L Aut, press

 

[UNIT] to select 

value (

X

 stands for AUT, OFF, ON), OFF stands for inactivate 

backlight function, ON stands for the backlight is always make function. AUTO 

stands for auto backlight. 

 

4. Press 

[ZERO]

 to return to weighing mode. 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Summary of Contents for AD-300i

Page 1: ... select ZERO track value 0 0d 0 5d 1 0d 1 5d 2 0d press ZERO key to enter ZERO track function setting display shows ZRT x xd press UNIT key to select ZERO track time 05 to 4 0 seconds press ZERO key to enter ZERO setting display shows RZD XXX press UNIT key to select ZERO value 0 8d 1 8d 2 8d 3 8d press ZERO key to enter FIL setting press UNIT key to select filter range amend 0 1 2 3 4 5 6 7 2 is ...

Page 2: ...rged static electricity could influence the weighing Discharge the static electricity As example the method is to use the electrification prevention spray to spray it to both side of weighing platform 1 SPECIFICATIONS Model number Capacity Division AD 300i 300g 0 01g AD 600i 600g 0 01g AD 3000i 3000g 0 1g AD 6000i 6000g 0 1g Pan Size φ 98mm or120 x 135mm or 170x190mm Net gross weight 2100 2800g Pa...

Page 3: ... requirement 3 Calibration by the weight When you selected the calibration weight and then place the calibration weight on platform Then press key the display returns to the weight value Now the calibration is complete 4 Press turn off the power press to turn on the power again place the weight on the platform to make sure whether weighing is correct if not repeat steps 2 4 5 Return to weighing mo...

Page 4: ... battery Weighing procedures 1 Press to turn on the balance When power is turned on all display segments appear on LCD for a few seconds and finally 0 will appear on the display 2 Select the weighing unit with Press to select a weighing unit e g g ct oz dwt Once the unit has been selected the selected unit will be displayed to the weight value 3 Start weighing If you do not use a container for wei...