Photoionization Detector Module PIM CPN
User Manual
PIM_CPN - User Manual (en) - 1.08.docx
Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com
Page 19 of 48
The procedure step by step:
•
check the appropriate gas flow
•
power on the module
•
wait for IDLE state (
Action ‘state’
)
•
enter MEAUSRE state (
Action ‘start’)
•
read the measurement result (
Action ‘values’ ) and module state (
Action
‘state’ )
Action
UART command
I2C data
start
pid.start
Control register set D00
lamp check
pid.lampcheck
Control register set D01
stop
pid.stop
Control register set D02
values
pid.values
Read measurement data registers
state
pid.state
Read Status register
error
pid.error
Read Error register
reboot
pid.reboot
Control register set D03
In case of ERROR state the error details can be readed
(Action ‘error’).