Installation
46
External emission light
Pin 22
EMISSION
24 V, max. 200 mA, fused by poly fuse.
Pin 22 carries 24 V when the shutter is open.
The connected emission warning light signals danger from laser radiation.
Laser ready
Pin 3
READY
24 V, max. 200 mA, fused by poly fuse.
This output produces a summation signal of all possible error states. The output is set
when all components function within normal parameters and the laser is ready for opera-
tion.
This signal can be processed both optically and electronically. This output is an essential
aid, particularly for use in production lines.
Inputs and outputs
Pin 5 to 12
IN1 to IN8
24 V, 3.5 mA
Control inputs for PLC connector
Pin 23 to 30
OUT1 to OUT8
24 V, max. 0.5 A
WARNING
DANGER OF INJURIES AND/OR RISK OF PROPERTY DAMAGE FROM
LASER RADIATION!
Wiring this output is specified for operation in accordance with a class
4 laser.
!