Ace ALCOSCAN Pro MED Operating Manual Download Page 17

 

 

 

 
 
 

Warning Display

 

                

 

FLOW 

If you see „FLOW“ on the display, the breath sample was not strong 
enough for a measurement. Please wait until you can see “BLOW” on 
the display again and then make sure that the breath is strong and 
steady during the sampling. 

Battery Change  

If you see this on the display then the battery power is low, please 

replace the batteries with 2 new ones. 

Service required 

Calibration 

This message is shown 6 months after the last calibration or by setting 

CAL required number. 

C.

 

Test No 

Number of tests after a calibration (memorized in the sensor module). 

T. total 

Number of all tests done with the device (memorized in the device). 

Sensor empty 

Check to connect 

system  

Error with sensor module, please replace the sensor with a new one. 

Over Temperature 

Keep to operating 

temperature range 

Error because of critical temperature during usage, the device must 
not be used under 5° C. 

System Error 1 

Service required 

device 

 

Condensor charging error, please contact ACE customer service.  

Sensor empty 

Check to  

connect system 

 

Sensor leakage, please contact ACE customer service. 

 
 
 
 
 
 
 
 
 
 
 
 
 

Summary of Contents for ALCOSCAN Pro MED

Page 1: ...tzerfreundliches digitales Alkoholtestger t mit elektrochemischem Sensor Es misst die Alkoholkonzentration in der Atemluft und berechnet daraus die Blutalkoholkonzentration Die Testergebnisse k nnen i...

Page 2: ...den Bezeichnung der einzelnen Bestandteile 1 LED Licht 2 IrDA Fenster 3 Digitale LCD Anzeige 4 W hltaste nach oben 5 Men W hltaste nach unten 6 Startknopf Auswahltaste 7 Schiebehebel zum ffnen der Bat...

Page 3: ...eben Sie den Hebel oben an der Ger ter ckseite nach rechts 2 Schieben Sie die Abdeckung nach unten und entfernen Sie diese vom Ger t 3 F hren Sie den d nnen Teil der Handschlaufe in die ffnung an der...

Page 4: ...nleitung bezieht enth lt im Auslieferungszustand zwei AA Alkaline Batterien Diese Batterien sind nicht dazu bestimmt wieder aufgeladen zu werden Der Versuch der Wiederaufladung kann Sach und Personens...

Page 5: ...M in das Men wechseln der Knopf befindet sich links oberhalb des Startknopfes Wenn Sie das Men wieder verlassen w rmt sich das Ger t automatisch weiter auf 5 Sobald Bereit erscheint pusten Sie f r zir...

Page 6: ...t 3 Entfernen Sie nun unbedingt die beiden Batterien s Abb 2 4 Dr cken Sie danach die beiden Klammern rechts und links am Sensormodul gleichzeitig zusammen s Abb 2 Mitte 5 Nehmen Sie das Sensormodul a...

Page 7: ...nden Sie das Ger t zur Kalibrierung ein oder erwerben Sie einen neuen Sensor Tests Kal Anzahl der durchgef hrten Tests nach der Kalibrierung die Anzahl wird im Sensormodul gespeichert T total Anzahl a...

Page 8: ...est count Test Gesamtanzahl d Date Time Datum Zeit e Configuration Konfiguration Einstellungen der Promillewerte bei der Testergebnis Ermittlung f Device info Ger teinformation II General setting Allg...

Page 9: ...t die Messgenauigkeit gepr ft wurde das Ger t speichert maximal 10 Werte c Calibration Kalibrierung hier kann der Hersteller in den Kalibrations Modus wechseln d Accuracy check Check der Messgenauigke...

Page 10: ...en Nebensch den oder Folgesch den wie unter anderem entgangene Gewinne und Nutzungsausfall die aus Verkauf Herstellung oder Nutzung der unter diesem Haftungsausschluss verkauften Produkte entstehen un...

Page 11: ...3 42 EEC for medically products It is a digital electrochemical alcohol testing device which measures the alcohol concentration in the breath and calculates the blood alcohol level from that For the A...

Page 12: ...lts It must only be used on basis of the instructions mentioned in this manual Label of individual components 1 IrDA window 2 Digital LCD display 3 Up button 4 Menu down button 5 Power select button 6...

Page 13: ...they are inside the device will cause damage to the device Fixing the hand strap 1 Push the button on the back of the device and remove the battery cover 2 Insert hand strap thorough the small hole in...

Page 14: ...ause damage to the environment or health Please return batteries rechargeable batteries to the store or to the buyback centre The return is free of charge und mandatory by law Please throw only deplet...

Page 15: ...ht is blinking above the display this green light blinks until you start your test 4 After taking your breath sample the LCD display shows Analyzing wait 5 The result is displayed with one beep sound...

Page 16: ...right side 2 Pull the battery cover down and remove the battery cover 3 Grab handles and pull down side like shown in the second picture 4 Take out the sensor module carefully 5 Place new sensor care...

Page 17: ...last calibration or by setting CAL required number C Test No Number of tests after a calibration memorized in the sensor module T total Number of all tests done with the device memorized in the device...

Page 18: ...f chooseable contens in the administer menu I Information g Last tests h Last maintenance i Total test count j Date Time k Configuration l Device info II General setting h Set language i Set date j Se...

Page 19: ...ration period s Tachograph connection to the tachometer of a vehicle this function is optional and not chooseable with this model t Measurement mode Normal or Fast default Normal to get exact test res...

Page 20: ...his warranty includes the electronics as well as the casing or the sensor without sampling components Components that are inevitably destroyed upon effectiveness such as fuses are not part of this war...

Page 21: ...er and the proper use maintenance and service according to the instructions of the accompanying product literature by the ACE Handels und Entwicklungs GmbH are mandatory prerequisites for this warrant...

Reviews: