2

 

 

STEP 3: START Up

 

Before  proceeding  with  start

-

up,  verify  the 

following items have been completed:

 

 

-

 All installation has been completed and verified.

 

 

-

 All wiring has been completed and verified.

 

 

-

 Power is present at the AVC6 and verified to spec.

 

 

-

 Exhaust system is operating with static pressure control.

 

 

-

 The configuration PC has Insight loaded and operational.

 

3a. Connect PC to FHM3

 

 

 

 

 

3b. 

On the AIRFLOW SETPOINT Gauge, Select 

MODE

 

 
 
 
 
 

 

 

3c. 

Verify the Mode of Operation is set to 

FUME HOOD MODE

 

 

 

 

 

 

 

 

 

 

The 

Calibration  Tools

  provided  for  Fume  Hood  Mode  are 

accessed through a series of buttons that are  positioned in the 

order  of  which  they  should be  completed  during  the  initial start 

up.

 

 

 

 

 

 

 

 

3d.

 

Select 

Configure  Fume  Hood  Type 

and  enter  the 

information required.

 

 

 

  

 

 
 

3e.

 

Select 

Calibrate Sash 

Input 

if sash sensors are employed and perform the sash 

calibration procedure.

 

 

  

 

 

 
 

 

3f.

 

Select 

Configure Control Setpoints 

and enter the setpoint 

information required for the application.

 

 

 

 

 

 

 

 

3g.

 

Select 

Configure Purge Mode 

and select the appropriate 

Purge configuration per the application requirements.

 

 
 

 
 
 

 

3h.

 

Select 

Configure Alarms

 and enter the alarm setpoint 

information required for the application.

 

 

 

 

 

 

 

 

 

4.    

After completing the above steps, verify the displayed face  

 

velocity correlates with the measured average face velocity.

 

Be sure the average face velocity field measurement is 

obtained using appropriate methods and instrumentation 

per industry standards and facility requirements.

 

 

5.    

If desired, the displayed face 

velocity can be correlated to   

the field measured face 

velocity by using the Face 

Velocity Measurement 

Adjust function, which can 

be accessed through the 

utilities Menu under Tools.

 

 

This document is a quick

-

start reference only and is not 

intended to be used exclusively as a comprehensive 

guide for the AVC6 Fume Hood Control System. For more 

details, refer to the AVC6 User Manual, submittal 

drawings and job specifications to obtain the required 

operating parameters. 

 

Plug  USB 

C

” 

Cable  from  PC  into 

the  FHM3  USB  Port  at  the  bottom 

of the Monitor.

 

Open  the  AVC  Insight 

UI Program.

 

AVC6 FUME HOOD CONTROL SYSTEM QUICK START GUIDE—

 