SENSOR REPLACEMENT 

In the event of a failed sensor (as indicated by the Failed Sensor 
Alarm), cleaning or replacement of the sensor is necessary. First, 
clean the sensor as directed on page 4. If an alarm is still heard, 
the sensor must be replaced. To replace the sensor: 

 

1.  Turn the power off. 

2.  Remove the sensor compartment cover by loosening the 

sensor compartment cover screw. 

3.  Remove  the 2 screws holding the sensor in place with a 1.5 

mm hex wrench. 

4.  Grasp the old sensor with your fingers near the screw holes 

and firmly lift out of the sensor compartment. 

5.  Firmly press the new sensor into the compartment, noting 

alignment of the sensor connector tab and gas ports of the 
sensor with the mating features in the instrument. 

6.  Firmly tighten sensor in place with 2 screws and 1.5 mm hex 

wrench. 

7.  Replace and fasten sensor compartment cover. 

 
Replacement Sensor Kit

  

 

Part Number 

ELS-7CO2

 