53
13.2
除霜溫度控制
- Defrosting temperature control
This function is refers to when the execution defrosts the movement, whether the precooling area
temperature does control assigning.
1.
有
-enable: Defrosted the conclusion temperature according to the hypothesis to carry out the PID
control.
2.
無
-Disable: Does not carry out the PID control.
13.3
試驗時間條件
- Testing time condition :H.M/M.S Choice
This function is refers to the experiment when the hypothesis or the execution, time condition
determination choice.
1. H.M: The experiment time setting or the execution/divide into the condition by the hour.
2.H.S: The experiment time setting or the execution take a minute/second as conditions.
13.4
保養項目
- Maintenance project
■
Maintenance setup:May register 1 0 places the inspection project name and the inspection time.
1. Maintenance name:
The single-byte character, may register 32 writing. (British digit or symbol)
The double-byte character, may register 15 writing. (Chinese character, British digit or
symbol)
2. Next time inspection time:
But scope of the hypothesis is 0 ~ 9999 time (day).
■
Maintenance control
:
Hand regulation control output output as well as D A M P E R input/output
data condition expression.
1. Control output operation: 0 ~ 100%. Heating chamber and low greenhouse simplex operation
2. DAMPER operation: Opens/closes the choice. (preheating area, test area and precooling
area simplex operation)
3. DAMPER input output indicated: Represents “ON” red, the black represents “OFF”.