8.  Maintenance

Always disconnect the power cord for cleaning.  

When cleaning the surfaces of the instrument, use a 
lightly dampened cloth with mild detergent if needed.  
Do not use any corrosive solutions that could damage 
plastic.

Use caution when cleaning near the housing vents.  
Do not allow liquid to enter the power supply.

9.  Troubleshooting

Problem

Causes

Measures

No display

Power not 
connected.

Fuse blown.

Check cord connection 
and power supply, 

Unplug and re-attach 
power cord.

Check and replace fuse 
at power inlet.

Voltage or 
Amperage setting 
not reached

Maximum wattage 
exceeded

Resistance of the load 
(gel tank buffer) is too 
low.  Change buffer 
concentration.

NO LOAD 
displayed

Connection wires 
not attached 
properly.  Buffer is 
not conductive.

Check wire 
connections.  Check 
buffer concentration.

Summary of Contents for myVolt Touch

Page 1: ...300V Power Supply Operating Manual a division of This product is for research only ...

Page 2: ...s that could damage plastic Use caution when cleaning near the housing vents Do not allow liquid to enter the power supply 9 Troubleshooting Problem Causes Measures No display Power not connected Fuse blown Check cord connection and power supply Unplug and re attach power cord Check and replace fuse at power inlet Voltage or Amperage setting not reached Maximum wattage exceeded Resistance of the l...

Page 3: ... sets of outputs that accept standard electrical leads with banana plugs and should only be used with electrophoresis tanks that incorporate sufficient safety features to prevent electric shock 1 7 Overview of Electrical Principles Definitions Voltage The amount of potential energy between two points in a circuit Current The time rate of flow of an electric charge measured in amperes Resistance A ...

Page 4: ...trol and Display 4 Pairs of Electrical Outputs Folding Front Feet Recesses for Stacking 2 Units AC Power Inlet and Fuse 6 4 Automatic Cross Over During an electrophoresis run the power supply will attempt to maintain the constant voltage or amperage setting depending on which parameter is set to CONSTANT If during a run the non constant parameter limit is met due to the resistance of the load the ...

Page 5: ...on constant parameter set as a limit Press the operation icon CONSTANT to switch between CONSTANT CURRENT and CONSTANT VOLTAGE running modes In constant voltage mode the power supply will output the set voltage and the mAmp current value will fluctuate depending on the resistance of the load buffer concentration in the attached electrophoresis tanks up to a maximum of the mAmp current setting that...

Page 6: ...ght strong light corrosive gas high magnetic fields heaters fires and other heat sources 4 5 6 Operating guide 6 1 Touch Screen Control Up Down Value Adjust Buttons Adjustable Parameters Operation Icon Buttons Status 4 Installation Place the power supply on a smooth level surface in an area where it will not contact liquids or chemicals The myVolt Power Supply features Dual Orientation technology ...

Reviews: