34
< PRINT >
It selects the printing mode of the A4 printer.
- Select
(PRINT) on SYSTEM SETUP screen and
press ‘SET’ button on touch pad.
- Pre-set: AUTO
- Choose one button on touch pad.
- Press ‘SET’ button on touch pad to save it.
- Return to SYSTEM SETUP screen by pressing ‘CLOSE’ button on touch pad.
< PRINT POSITION >
It adjusts the printing position in the direction of U-D
(up-down) and L-R (left-right) to fit to the pre-formatted
result sheet.
- Select
(PRINT POSITION) on SYSTEM SETUP
screen and press ‘SET’ on touch pad.
- Pre-set: 00 for U-D and 00 for L-R
- Range: 99 for U-D and 99 for L-R
- Choose U-D by pressing ‘▼, ▲’ on touch pad. Choose L-R by pressing ‘
▶
,
◀
’ on touch pad.
- Pressing
▶
button moves print position down or right.
- Pressing
◀
button moves print position up or left.
- Every single press moves print position by about 0.2 mm.
- Press ‘SET’ button on touch pad to save it.
- Return to SYSTEM SETUP screen by pressing ‘CLOSE’ button on touch pad.
Note
1. L-R (left-right): - is moving to the left and + is moving to the right.
U-D (up-down): - is moving up and + is moving down.
2. The print position needs to be separately set on the result sheet for Body
composition analysis and Segmental analysis each.
Press “BCA” and adjust the print position of Body composition analysis
result sheet and then press ’SET’ button to save the setting.
Press “SEG” and adjust the print position of Segmental analysis result
sheet and then press ’SET’ button to save the setting.
It is possible to adjust the print position of the result sheet for children if you
press the CHD button.
3. Print test
When you press “Print” button, you can print the test page of BCA, SEG or
CHD.
Summary of Contents for BC720