40
AcuDC 200 Series
The third screen: Shunt / Hall Effect Sensor
current input full range setting. The value
input is the full range current value for the
Shunt or Hall Effect Sensor. AcuDC 200 series
support 1~9999A shunt. The value range in
this screen is 1~9999.
The fourth screen: Shunt / Hall Effect Sensor
output voltage setting. The value input is the
full range output voltage value for the Shunt
or Hall Effect Sensor. For instance, 100mV
shunt, enter 100; 5V Hall Effect Sensor, enter
50; 20mA Hall Effect Sensor, enter 20.
AcuDC 200 series support 50~100mV arbitury voltage shunt. If it exceeds this range,
the value will be reset to default 100. Our recommendation for shunt is 100mV
shunt, it is better for accuracy.
The value here for shunt, the range is 50~100; for volage Hall Effect Sensor, the
range is 1~100; for current Hall Effect Sensor, the range is 4-100. 4~XmA and 0~XmA
types are not exchangable.
A
Fig 3.10 Shunt current input full range setting
Fig 3.11 Shunt/Hall Effect Sensor output setting