EXC-400 MICROSCOPE SERIES
ACCU-SCOPE
®
73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com
10
DETAILED ASSEMBLY
(continued)
Installing the Stage -
(Fig. 6 & 7)
Lower the condenser to its lowest position by turning the
condenser carrier knob counterclockwise
①.
Remove protective tape and bag from stage and
specimen holder
②
. (The stage comes with the
specimen holder pre-mounted to the top of the stage).
Loosen the thumb screw
③
on the front of the stage.
Position the X-Y movement knobs
④
to the right and
align the circular mount on the bottom of the stage with
the circular bracket
⑤
on the condenser/stage carrier
and set the stage in place so that it is centered over the
condenser below.
Tighten the thumb screw
③
on the front of the stage.
Raise the condenser carrier by turning the knob
clockwise.
Figure 7 shows the stage in place.
Installing the Viewing Head -
(Fig. 8)
Using the 5mm Hex wrench, loosen the lock screw
①.
Position the viewing head
②
above the dovetail opening
③
as shown and with it tilted slightly down on the right,
slide the dovetail under the notches in the dovetail hole
and set it into place with the two eyepiece tubes
④
facing forward.
Retighten the lock screw
①.
Fig. 6
Fig. 7
①
④
①
②
③
③
②
③
④
Fig. 8
⑤