background image

EXC-100 MICROSCOPE SERIES 

 

73 Mall Drive, Commack, NY 11725 • 631

-864-

1000 • 

www.accu-scope.com 

12 

 

ADJUSTMENT & OPERATION

 (continued)

 

Using Binocular and Trinocular Heads 

 

Fig. 13 

 

Adjusting Interpupillary Distance 

1.  When observing with two eyes, hold the base of the eyetubes and 

rotate them around the axis until there is only one field of view.   

”mark 

 on the eyetube base points to the scale

 of 

interpupillary indication, means the value of interpupillary distance 

(see Fig. 13). 

The adjustable range is 50~75mm. 

 

  Remember your interpupillary distance for fast setting in future 

operation.    Your interpupillary distance can be used with other 

microscopes, too. 

 

Fig. 14 

 

Fig. 14 

 

Assembling and Using the Trinocular Port and Camera 

(trinocular model only) 

1.  Loosen the lock screw

 of trinocular head and remove the 

dustcover

 (See Fig. 14). 

2.  Remove the dustcovers from the camera adapter assembly

.   

Thread the top of the camera adapter into the C-mount of the 

camera.    Place the assembly into the trinocular head (flange side 

towards microscope; camera end points up) as shown in the figure 

and screw down the lock screw

 

3.  For binocular observation, focus through the eyepieces until the 

image is sharp.    Observe the image from the camera. If the image is 

out of focus, rotate the focus ring on the camera adapter

 to adjust 

the focus in the camera image until it is sharp.     

4.  If the image from the camera is rotated compared to the eyepieces, 

loosen the lock screw

 and rotate the camera until the images have 

the same alignment.    Retighten the lock screw. 

 

Summary of Contents for EXC-100 Series

Page 1: ...73 Mall Drive Commack NY 11725 631 864 1000 www accu scope com info accu scope com MANUAL EXC 100 MICROSCOPE SERIES...

Page 2: ...G THE 100X OIL IMMERSION OBJECTIVE 11 A MOVE THE EYEPIECE TO EXAMINE THE AIR BUBBLE OPEN THE APERTURE DIAPHRAGM AND FIELD DIAPHRAGM FULLY AND OBSERVE THE EDGE OF THE OBJECTIVE FROM THE TUBE IT SEEMS R...

Page 3: ...t including eyepieces objectives or focusing assembly 2 Keep the instrument clean remove dirt and debris regularly Accumulated dirt on metal surfaces should be cleaned with a damp cloth More persisten...

Page 4: ...icroscope from the Styrofoam container by its arm and base and place the microscope on a flat vibration free surface Operation Notice Fig 1 1 As the microscope is a high precision instrument always op...

Page 5: ...SERIES 73 Mall Drive Commack NY 11725 631 864 1000 www accu scope com 5 COMPONENTS DIAGRAM Objective Stage Illuminator Fine Focusing Knob Coarse Focusing Knob Focus Tension Adjustment Ring Frame Binoc...

Page 6: ...Y 11725 631 864 1000 www accu scope com 6 Eyepiece Trinocular Head Fine Focusing Knob Coarse Focusing Knob Frame Objective Stage Condenser Lock Screw Indicator Light Light Adjustment Knob Nosepiece Ve...

Page 7: ...ians at our New York facility prior to shipment Should you need to disassemble assemble your microscope in the future please follow the instructions outlined below When assembling the microscope make...

Page 8: ...objective nosepiece until it sounds ka da to make sure the desired objective is in the center of optical path Fig 4 Installing the Eyepiece s 1 Remove the cover from the eyepiece tube 2 Insert the ey...

Page 9: ...connected to ground ADJUSTMENT OPERATION Fig 7 Illumination 1 Turn on the power by rotating the light adjustment knob and adjust light intensity by rotating the knob until the illumination is comforta...

Page 10: ...focus ring located just below the right eyepiece until the image seen in the right eyepiece matches the left eyepiece The focus travel locking screw sets the maximum height that the stage can move wh...

Page 11: ...ted into the light path Fig 12 Using the 100x Oil Immersion Objective 1 Use the 4X objective to focus the specimen 2 Place a drop of oil on the specimen observed see Fig 12 3 Rotate the nosepiece coun...

Page 12: ...ssembling and Using the Trinocular Port and Camera trinocular model only 1 Loosen the lock screw of trinocular head and remove the dustcover See Fig 14 2 Remove the dustcovers from the camera adapter...

Page 13: ...and allow the clip to slowly close onto the slide and clamp the slide firmly see Fig 15 Do not let go of the lever as the clamp will close too fast This may result in slide breakage and flying glass...

Page 14: ...wer the condenser To lower the condenser grasp the metal ring on the condenser and rotate counterclockwise to the left After changing the position of the darkfield stop raise the condenser back to the...

Page 15: ...ion by swinging the objective correctly into the optical path Dirt or dust on the view field Dirt or dust on the lens eyepiece condenser objective collector lens or specimen Clean the lens Poor image...

Page 16: ...ating the intensity control dial and or iris diaphragm Open to the proper setting Position the condenser slightly lower than the upper limit MECHANICAL PROBLEMS Image will not focus with high power ob...

Page 17: ...e of the condenser is too low Raise the condenser base Slide does not move smoothly The slide is not placed correctly The movable specimen holder is not mounted properly Place the slide in the holder...

Page 18: ...Commack NY 11725 631 864 1000 www accu scope com 18 enough The light adjustment knob is set too low The light adjustment knob is not working correctly one Adjust the light intensity correctly Replace...

Page 19: ...nship for a period of five years from the date of invoice to the original end user purchaser The LED lamp is warranted for a period of one year from the date of invoice to the original end user purcha...

Reviews: