3
14
The ACCU-CHEK FastClix Lancing Device
Inserting a Lancet Drum
You must first load the lancet drum into the lancing device to get it ready for use.
1
Remove the cap.
Insert a new lancet drum,
white end first, until it
clicks firmly into place.
2
Slide the cap on until it
stops by aligning the
notch on the cap with the
notch on the lancing
device.
3
You are now ready to use
the first lancet.
The lancet counter shows
a number 6, meaning you
have six new lancets
remaining.
WArNINg
You must not insert the lancet drum into the lancing device and simultaneously press the release button or
hold the lancing device with the release button resting on a surface such as a table top. This could release a
lancet and inadvertently cause injury.
NOTE
Once the lancet drum is inserted into the lancing device, do not remove the lancet drum until completely
used. The lancet drum cannot be reused once it has been removed from the lancing device.
51742_06333630003.indb 14
1/22/13 1:17 PM