This message/
symbol appears ...
and means:
The test is in progress.
MEMO
MEMO
indicates a value saved in memory.
MEMO
No results have been saved.
The meter cannot calculate the average
value
as memory only contains results that are
not considered in the calculation, e.g.
quality control results,
as you have not entered the time and
date,
as you have set the date to a time in
the past.
This message/
symbol appears ...
and means:
CTRL
You are requested to run a quality control test.
CTRL
This is/was a quality control test.
The temperature during testing is at the limits of
the permitted range (b5 °C and +10 °C
or b40 °C and +45 °C).
Do not use the result as a basis for therapeutic
decisions.
Turn the meter off. Let the meter slowly cool
down or warm up. Turn the meter on and repeat
the test. The optimum measuring temperature is
b10 °C and +40 °C.
MEMO
The result in memory was obtained outside the
permitted temperature range.
Do not use the result as a basis for therapeutic
decisions.
9.1 Messages and symbols