51
Meter Maintenance and Troubleshooting
8
(The error code and
Err
will alternate on the display.)
An electronic error occurred.
Remove the batteries, press and hold the
power/set button for at least 2 seconds, and
reinsert the batteries. Perform a blood glucose or
control test.
(The error code and
Err
will alternate on the display.)
The temperature is above or below the proper
range for the system.
Refer to the test strip package insert for system
operating conditions. Move to an area with the
appropriate conditions and repeat the blood
glucose or control test. Do not artificially heat or
cool the meter.
(The error code and
Err
will alternate on the display.)
Not enough blood or control solution was drawn
into the test strip for measurement or was
applied after the test had started.
Discard the test strip and repeat the blood
glucose or control test.
(The error code and
Err
will alternate on the display.)
Blood or control solution was applied to the test
strip before the flashing drop symbol appeared
on the display.
Discard the test strip and repeat the blood
glucose or control test.
1080_Guide Me_MANUAL_MASTER.indb 51
3/5/2018 1:28:03 PM
Summary of Contents for Guide Me
Page 2: ...1080_Guide Me_MANUAL_MASTER indb 2 3 5 2018 1 27 39 PM ...
Page 4: ...1080_Guide Me_MANUAL_MASTER indb 4 3 5 2018 1 27 39 PM ...
Page 32: ...30 Control Tests 4 1080_Guide Me_MANUAL_MASTER indb 30 3 5 2018 1 27 53 PM ...
Page 36: ...34 Review Your Data 5 1080_Guide Me_MANUAL_MASTER indb 34 3 5 2018 1 27 55 PM ...
Page 64: ...62 Technical Information 9 1080_Guide Me_MANUAL_MASTER indb 62 3 5 2018 1 28 06 PM ...
Page 68: ...66 License and Warranty 10 1080_Guide Me_MANUAL_MASTER indb 66 3 5 2018 1 28 06 PM ...
Page 70: ...68 Index 1080_Guide Me_MANUAL_MASTER indb 68 3 5 2018 1 28 06 PM ...
Page 71: ...1080_Guide Me_MANUAL_MASTER indb 69 3 5 2018 1 28 06 PM ...