Accsense
VersaLog Data Loggers User’s Manual
Page 54
The following fields are for editing:
Description:
Description specifies the information about the logging session with a maximum of 30
characters. It will be the default Title section of the plot in the plot view.
Sampling Interval:
This field specifies the time span the logger will wait between two measurements sampling.
Valid settings are:
Sampling Interval for Fast
Sampling Mode
Sampling Interval for Normal Sampling Mode
20, 30,…90 milliseconds
5 seconds to 9 seconds in 1-second increment
100, 200,… 900 milliseconds
10 seconds to 50 seconds in 10-second increment
1, 2,3,4 seconds
1 minute to 59 minutes with 1-minute increment
1 hour to 12 hour with 1-hour increment
If the sampling interval is four seconds or faster the logger must be powered by an external
power supply during the logging period.
Making changes to the Sampling Interval will affect Total Time Span fields.
On-Board LED
Check this field to enable the on-board status LED. If the LED is enabled it will flash each
time when it samples data to indicate:
1.
The logging is active if the LED flashes in green.
2.
The logger is in alarm state if the LED flashes in red.
3.
The battery will die soon if the LED flashes in amber.
If you do not need LED indication, you can uncheck this field to increase the battery life.
When Memory Full
If you want the logger to stop logging when the memory is full select
Stop logging
. If you
want the logger to continue logging and overwrite the oldest data with the new data (FIFO),
you choose
Continue logging
.