background image

 

 

21 

 

 

 

Guidance and manufacturer’s declaration

-electromagnetic immunity 

 

The chair is intended for use in the electromagnetic environment specified below. 

The customer or the user of the chair should assure that is used in such and environment. 

Immunity test 

IEC 60601 test level 

Compliance 
level 

Electromagnetic environment-
guidance 

Conducted RF 

IEC 61000-4-6 

 

 

Radiated RF 

IEC 61000-4-3 

3 Vrms 

150 KHz to 80 MHz 

 

 

3 V/m 

80MHz to 2,5 GHz 

3 Vrms 

 

 

3 V/m 

Portable and mobile RF 
communications equipment should be 
used no closer to any part of the chair 
including cables, than the 
recommended separation distance 
calculated from the equation 
applicable to the frequency of the 
transmitter. 

Recommended separation distance: 

d = 1,2 √P

 

d = 1,2 √P 80MHz to 800 MHz

 

d = 2,3 √P 800MHz to 2,5 GHz

 

Where P is the maximum output power 
rating of the transmitter in watts (W) 
according to the transmitter 
manufacturer and d is the 
recommended separation distance in 
metres (m). 

Field strengths from fixed RF transmitters, 
as determined by an electromagnetic 
site survey, a should be less than the 
compliance level in each frequency 
range Interference may occur in the 
vicinity of equipment marked with the 
following symbol: 

 

NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies. 

NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by 
absorption and reflection from structures, objects and people. 

 

 

 

 

 

 

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) 
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV 
broadcast cannot be predicted theoretically with accuracy. To assess the 
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey 
should be considered. If the measured field strength in the location in which the chair is 
used exceeds the applicable RF compliance level above, the chairshould be observed to 
verify normal operation. If abnormal performance is observed, additional measures may be 
necessary, such as re-orienting or relocating the chair. 

 

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. 

Summary of Contents for Configura Advance Chair

Page 1: ...Configura Advance Chair Instruction Manual IFU SC1 001EN Rev 04 May 2022...

Page 2: ...ified as a Class 1 Medical Device in accordance with the Medical Devices Regulation 2002 as amended and the Medical Device Regulation 2017 745 Notice to User If a serious incident occurs in relation t...

Page 3: ...ents may be used otherwise Accora cannot guarantee against any damage that might occur 14 Inappropriate routing of accessory cables e g cushion pump cable could lead to dangerous electrical hazards if...

Page 4: ...ng homes or convalescent units The Configura Advance chair may also be placed in private homes where there is unlikely to be 24 hour supervision by clinical staff but where the patient or caregiver is...

Page 5: ...specification Mains electrical system Description Value Supply voltage 100 240VAC Supply frequency 50 60Hz Max supply current 3 9 Amps Degree of protection against liquid ingress IPX4 Degree of protec...

Page 6: ...and hold in place Align the mounting hole at the bottom of the gas strut Fig 3 Item 2 and the mounting hole at the bottom of the backrest lockout slide Fig 3 Item 3 with the clevis at the back of the...

Page 7: ...e other side 10 Secure the backrest cushions to the backrest with the Velcro flaps Each cushion is attached to the backrest with the Velcro flap at the top of the cushion The top cushion has a double...

Page 8: ...djusted from 330mm 13 to 533mm 21 The height is adjusted using two plungers Chair Width adjustment 1 Make sure a patient is not in the chair 2 Make sure the brakes on all 3 locking castors are locked...

Page 9: ...ir 2 Undo the hand screw and sidepiece locking plunger Figure 8 3 Pull the sidepiece from the chair seat frame undo the screws Figure 15 from the fixing bars Figure 15 4 Locate the fixing bars to the...

Page 10: ...that the plungers that secure the footplate Fig 18 Items 1 2 are in the closed position Fig 19 2 Take the footplate Fig 21 and insert the two tapered spigots Fig 21 Items 1 2 into the two sockets eith...

Page 11: ...is used to adjust this angle 3 Manual Forward Tilt The chair can tilt forwards from 7 degrees normal tilt to 10 degrees of forward tilt with lock bolts removal 4 Manual Leg rest The leg rest can be ad...

Page 12: ...e 3 Release the lever to lock the tilt in space angle Manual Forward Tilt Adjustment This function enables some patients to exit the chair more easily however it must only be used following a risk ass...

Page 13: ...est lockout slide Fig 29 Item 1 3 Release the backrest lockout slide by lifting the locking lever away from the backrest lockout slide The backrest can now be adjusted using the backrest lever 4 Pull...

Page 14: ...f the patient 1 Remove the footplate See section 9 2 Make sure the brakes on all 3 locking castors are locked on 3 Ensure it is safe to move the chair into the forward tilt position 4 Press buttons 4...

Page 15: ...the charging lead is stretched the two sections may come apart from each other Move the chair nearer to the mains socket and reconnect the two sections of the charging lead We recommend that the batte...

Page 16: ...he lever will flip up 3 Push the chair forward moving slightly from side to side to ensure 4 To unlock the castor swivel press on the top of the green lever to bring it back horizontal Directional loc...

Page 17: ...o not use detergents that could alter the structure or behaviour of the plastics petrol etc Clean by wiping with a damp cloth Alternatively use Microfibre cleaning cloth without any soap For extreme s...

Page 18: ...user or another person The following checks must be carried out 1 Does the chair operate as per its intended purpose without unexpected noise or motion 2 Are there any signs of abuse or excessive wear...

Page 19: ...SPR 11 Group 1 The chair uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions...

Page 20: ...s to earth 1kV differential mode Not applicable Mains power quality should be that of a typical commercial or hospital environment Voltage Dips short interruptions and voltage variations on power sup...

Page 21: ...nsmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the followi...

Page 22: ...ance according to frequency of transmitter M 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 1...

Page 23: ...inst liquid ingress Do not dispose of in household waste Degree of protection against electric shock Class II Double Insulated For indoor use only Medical Device in accordance with EU Medical Device R...

Page 24: ...24 IFU SC1 001EN REV 04 Accora Barrington Road Orwell Cambridge SG8 5QP United Kingdom T 44 0 1223 206100 info accora care www accora care...

Reviews: