15 

12.

 

Battery operation and charging 

 

The Configura Advance electric version has a 
rechargeable battery fitted to power the electric 
functions.   
When a handset button is pressed, the relevant 
function will operate. 
If a warning beep sounds when any handset button is 
pressed, the battery may be low on charge and must 
be recharged as soon as possible.  
Please see Section 16 Troubleshooting if the chair does 
not operate as expected. 
 

 

 
The charging lead is in two sections.  The short section 
plugs into the control box, and the long section plugs 
into a mains socket. 
The two sections are connected by an inline plug and 
socket. To charge the Configura Advance: 
 
1.

 

Ensure the round connector is plugged into the 
control box on the back of the chair. 

2.

 

Ensure the two sections of the charging lead are 
connected securely. 

3.

 

Plug the mains plug into a mains socket. 

 

Ensure that the charging lead does not create a trip 
hazard, and do not move the chair while the battery is 
being charged. 
If the chair is moved inadvertently and the charging 
lead is stretched, the two sections may come apart 
from each other.  Move the chair nearer to the mains 
socket and reconnect the two sections of the charging 
lead. 
We recommend that the battery is charged weekly.  If 
the chair functions are used a lot, then it should be 
charged daily. 
When charging is complete, disconnect the long 
charging lead at the inline connector and store the 
lead away from the chair. 
 
The battery box and control box are on the rear of the 
backrest.  There are two warning lights as shown in 
Figure 32. 
 
 

Control Box Warning Light 

See Figure 32 Item 1. 
This warning light will show the following: 
Solid green: 

Chair plugged into mains electricity. 

No light:  

Normal operation. 

Solid orange: 

Handset button pressed. 

 

Battery Box Warning Light 

See Figure 32 Item 2. 
This warning light will show the following: 
Solid yellow: 

Battery charging 

No light:  

Normal operation 

Flashing yellow:  Error while charging 
 
If there is an error while charging, check that all 
charging lead connections are secure.  Contact 
Accora if problem persists. 

 

 
 

Battery Storage 

Follow the directions in Section 4 for storage conditions.  
The battery must be recharged at least every 12 
months when in storage. 
 

 

 

Figure 31 

 

Figure 32 

Summary of Contents for CHAIR-0-SC1-030

Page 1: ...Configura Advance Chair Instruction Manual IFU SC1 001EN Rev 04 May 2022...

Page 2: ...ified as a Class 1 Medical Device in accordance with the Medical Devices Regulation 2002 as amended and the Medical Device Regulation 2017 745 Notice to User If a serious incident occurs in relation t...

Page 3: ...ents may be used otherwise Accora cannot guarantee against any damage that might occur 14 Inappropriate routing of accessory cables e g cushion pump cable could lead to dangerous electrical hazards if...

Page 4: ...ng homes or convalescent units The Configura Advance chair may also be placed in private homes where there is unlikely to be 24 hour supervision by clinical staff but where the patient or caregiver is...

Page 5: ...specification Mains electrical system Description Value Supply voltage 100 240VAC Supply frequency 50 60Hz Max supply current 3 9 Amps Degree of protection against liquid ingress IPX4 Degree of protec...

Page 6: ...and hold in place Align the mounting hole at the bottom of the gas strut Fig 3 Item 2 and the mounting hole at the bottom of the backrest lockout slide Fig 3 Item 3 with the clevis at the back of the...

Page 7: ...e other side 10 Secure the backrest cushions to the backrest with the Velcro flaps Each cushion is attached to the backrest with the Velcro flap at the top of the cushion The top cushion has a double...

Page 8: ...djusted from 330mm 13 to 533mm 21 The height is adjusted using two plungers Chair Width adjustment 1 Make sure a patient is not in the chair 2 Make sure the brakes on all 3 locking castors are locked...

Page 9: ...ir 2 Undo the hand screw and sidepiece locking plunger Figure 8 3 Pull the sidepiece from the chair seat frame undo the screws Figure 15 from the fixing bars Figure 15 4 Locate the fixing bars to the...

Page 10: ...that the plungers that secure the footplate Fig 18 Items 1 2 are in the closed position Fig 19 2 Take the footplate Fig 21 and insert the two tapered spigots Fig 21 Items 1 2 into the two sockets eith...

Page 11: ...is used to adjust this angle 3 Manual Forward Tilt The chair can tilt forwards from 7 degrees normal tilt to 10 degrees of forward tilt with lock bolts removal 4 Manual Leg rest The leg rest can be ad...

Page 12: ...e 3 Release the lever to lock the tilt in space angle Manual Forward Tilt Adjustment This function enables some patients to exit the chair more easily however it must only be used following a risk ass...

Page 13: ...est lockout slide Fig 29 Item 1 3 Release the backrest lockout slide by lifting the locking lever away from the backrest lockout slide The backrest can now be adjusted using the backrest lever 4 Pull...

Page 14: ...f the patient 1 Remove the footplate See section 9 2 Make sure the brakes on all 3 locking castors are locked on 3 Ensure it is safe to move the chair into the forward tilt position 4 Press buttons 4...

Page 15: ...the charging lead is stretched the two sections may come apart from each other Move the chair nearer to the mains socket and reconnect the two sections of the charging lead We recommend that the batte...

Page 16: ...he lever will flip up 3 Push the chair forward moving slightly from side to side to ensure 4 To unlock the castor swivel press on the top of the green lever to bring it back horizontal Directional loc...

Page 17: ...o not use detergents that could alter the structure or behaviour of the plastics petrol etc Clean by wiping with a damp cloth Alternatively use Microfibre cleaning cloth without any soap For extreme s...

Page 18: ...user or another person The following checks must be carried out 1 Does the chair operate as per its intended purpose without unexpected noise or motion 2 Are there any signs of abuse or excessive wear...

Page 19: ...SPR 11 Group 1 The chair uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions...

Page 20: ...s to earth 1kV differential mode Not applicable Mains power quality should be that of a typical commercial or hospital environment Voltage Dips short interruptions and voltage variations on power sup...

Page 21: ...nsmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the followi...

Page 22: ...ance according to frequency of transmitter M 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 1...

Page 23: ...inst liquid ingress Do not dispose of in household waste Degree of protection against electric shock Class II Double Insulated For indoor use only Medical Device in accordance with EU Medical Device R...

Page 24: ...24 IFU SC1 001EN REV 04 Accora Barrington Road Orwell Cambridge SG8 5QP United Kingdom T 44 0 1223 206100 info accora care www accora care...

Reviews: