Accelerated Care Plus OMNISOUND 3000C PRO User Manual Download Page 9

6                                                                                                                                      OMNISOUND

®

 3000C PRO USER MANUAL 

 

 

 

COPYRIGHT © 2001 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED 

 

Failure  to  follow  the  manufacturer’s  prescribed  maintenance  for  this  device  may  lead  to  device  failure  and 
transient  or  unreliable  performance.    State  and  federal  survey  and  JCAHO  require  all  equipment  to  be 
maintained and calibrated according to the manufacturer recommended schedules. 

 

A potential electric shock hazard exists once the device outer casing has been in part, or fully, removed. Only 
qualified service personnel should perform Service and repairs. Warranty will be voided if the outer casing has 
been removed or tampered with. 

 

Inspect and cleanse the skin prior to application.  Following treatment check the skin for evidence of irritation 
or  burns,  and  if  present,  treat  as  appropriate.    If  the  patient  has,  or  complains  of,  skin  irritation  following 
treatment;  shorten  the  treatment  time  on  the  next  treatment  session,  or  use  an  alternative  type  of  therapy  or 
applicator placement. 

 

Gradually increase the output intensity/power to required dose or patient tolerance while monitoring the device display. 

 

Caution  should  be  taken  with  patient  exhibiting  psychological  or  physical  hypersensitivity  to  the  therapeutic 
treatment. Several attempts should be made to place them at ease so that their confidence and cooperation can 
be gained during the treatment. 

 

The treatment area should be checked from time to time, and if there is evidence of, or if the patient complains 
of, pain or overheating during treatment, adjust the output downward until it is tolerated by the patient.  If the 
patient continues to complain of pain or overheating, discontinue the treatment and shorten the treatment time 
on the next treatment session, or use an alternative type of therapy or applicator placement. 

 

Do not apply treatment directly over/under hot or cold packs.  Caution is recommended when treatment follows 
the application of hot or cold therapy, which may alter the patient’s sensation.  Application of thermal agents 
over  areas  of  impaired  circulation  should  be  performed  with  caution  as  the  circulation  may  be  insufficient  to 
heat or cool the tissue, altering the patient’s perception of warmth and pain, and burns or tissue necrosis may 
result from subsequent treatment. 

 

Treatment  should  not  follow  the  application  of  medicated  patches,  salves,  or  creams.  The  presence  of  electrical 
stimulation may be altered by the presence of these materials on the patient’s skin. Some medications can alter the 
patient’s  sensation.  Heat  can  also  increase  the  absorption  of  medication  and  may  be  contraindicated.  If  there  is  a 
medical necessity to perform such treatments, these patients should be monitored diligently during application. 

 

Caution should be used when applying thermal ultrasound over areas of body where circulation is impaired, or 
which lack normal sensation. Absent or diminished sensation should be avoided or, if unavoidable, treated with 
caution.  Establishment of acceptable intensity levels for desensitized areas may be related to the intensity levels 
tolerated on normal skin in opposite or related body parts. 

 

Do  not  connect  this  device  to  any  electro  medical  equipment  for  combination  therapy  except  the  Omnistim

®

 

family of electro stimulators. 

 

Do  not  immerse  any  part  of  the  operator’s  body  into  the  water  bath  during  underwater  techniques  since  the 
long-term biophysical effects of ultrasonic energy exposure have not been evaluated. 

 

Over acute skin conditions such as eczema, dermatitis, etc.  

 

Summary of Contents for OMNISOUND 3000C PRO

Page 1: ......

Page 2: ...not be used for any other purpose Any additional copies of the Manual shall be ordered from ACP No changes or modifications shall be made to the Manual without prior review and written authorization...

Page 3: ......

Page 4: ...Ultrasound Pulsed Shortwave Diathermy devices and advanced Therapeutic Exercise Systems Our MEGAPULSE NEUROPROBE OMNISTIM OMNISOUND OMNIBAND OMNITEST OMNICYCLE and OMNIVR represent the most recent wor...

Page 5: ...ocuments Caution electrical precautions Remote Control connection CE mark of confidence compliant to MDD 93 42 EEC Connection Electrode Cable IP2X Classification against intrusion of water dust accide...

Page 6: ...CIPLES OF USE 15 Definitions 15 Universal Precautions Body Substance Isolation 15 Cleaning Disinfecting of the Omnisound 3000C Pro 15 Cleaning and Low Level Disinfection 15 Intermediate Level Disinfec...

Page 7: ...crylate or high density polyethylene Arthroplasties the effect on bony ingrowth arthroplasties is not well defined for this reason the most prudent course is avoiding ultrasonic therapy over these are...

Page 8: ...or neck triangle including stellate ganglion vagus nerve or laryngeal or pharyngeal muscle Particular care should be taken for patients with a known sensitivity to the carotid sinus reflex as carotid...

Page 9: ...ontinue the treatment and shorten the treatment time on the next treatment session or use an alternative type of therapy or applicator placement Do not apply treatment directly over under hot or cold...

Page 10: ...at make the Omnisound 3000C Pro truly the future of ultrasound therapy Controlled Depth of Penetration The Omnisound 3000C Pro enables you to treat soft tissue injuries superficially with 3 MHz and in...

Page 11: ...FACTOR buttons are used to select the ultrasound output Continuous 100 for thermal applications Pulsed 5 10 20 25 33 or 50 for sub thermal applications NOTE Duty factor starts at continuous and goes...

Page 12: ...display in watts power or watts cm intensity Q The cooling fan provides airflow through the unit when the unit is operational R The ELECTROSTIM INPUT PORT is used to connect the shrouded pin to banan...

Page 13: ...off the system and connect the transducer to the Main output connector If both transducers are connected the main transducer will be selected 2 Omnisound 3000C Pro transducers contain their calibrati...

Page 14: ...rapidly if the output is switched ON and the transducer is not in contact with the skin Power Supply Operation Omnisound 3000C Pro is a line power only device It is not intended for use with batteries...

Page 15: ...ucers are fully waterproof and may be used with underwater techniques It is recommended that the handle is held by the operator without direct contact to the water to avoid exposure The Omnisound 3000...

Page 16: ...t a lower output is recommended Avoid using maximum power levels For example a typical comfortable output should be 0 5 1 0 W cm2 4 Apply coupling medium plentifully to the area to be treated As the s...

Page 17: ...s the active electrode Attach the self adhesive electrode to the patient s body Take the negative black tip pin and insert it into the shrouded pin to banana adapter Item No 2503036 Insert the adapter...

Page 18: ...inal secretions is likely to occur Health care workers with exudating lesions or weeping dermatitis should refrain from all direct patient care and from handling patient care equipment until the condi...

Page 19: ...creased risk of patient cross contamination After each use clean common contact surfaces such as control panel and ultrasound transducer with ACP germicidal wipes With a second ACP germicidal cloth wi...

Page 20: ...e Omnisound 3000C Prounit 4 Select and cut with clean scissors a 2 foot length of plastic sleeve and slide it over the ultrasound handle and cable 5 Apply ultrasound gel into plastic transducer cover...

Page 21: ...ing in pulsed or continuous modes at 1 0 MHz or 3 0 MHz for therapeutic applications Theory of Operation Ultrasonic therapy devices generally consist of a generator and a transducer The generator prod...

Page 22: ...ted that the effective radiating area does not generally cover the entire surface area of the transducer effective radiating surface and varies with the crystal frequency when operated at different cr...

Page 23: ...20 OMNISOUND 3000C PRO USER MANUAL COPYRIGHT 2001 2017 ACCELERATED CARE PLUS CORP ALL RIGHTS RESERVED...

Page 24: ...rally produce beams of perfectly uniform intensity over the surface of the transducer Thus the measurement represents an average for the effective radiating surface area Beam Divergence The beam diver...

Page 25: ...e BNR was 7 1 the maximum intensity in the tissue would equal 7 watts cm2 at certain portions of the field This should be kept in mind when determining output dosage It is also important to maintain c...

Page 26: ...inuous overheating of the transducer Ultrasound Transmission and Absorption When ultrasonic energy is applied to the tissue it is absorbed in varying degrees This absorption of energy can increase tis...

Page 27: ...supply adequate couplant apply the transducer to the patient and increase the power As you do so you will notice that the timer will change from 15 minutes to a lower value This will only occur if th...

Page 28: ...ity contact the Service Center TRANSDUCER NOT PRESENT message is displayed on the unit Transducer may not be connected properly Check if transducer is plugged in correctly If not plug in transducer an...

Page 29: ...andard 21 CFR 1050 10 instituted by the Food and Drug Administration Periodic calibration of treatment parameters having a direct influence on the physiological effect of ultrasound on tissue i e ultr...

Page 30: ...n Rate 20 20 1 1 100 continuous continuous 2 1 50 2 ms 250Hz 3 1 33 2 ms 165Hz 4 1 25 2 ms 125Hz 5 1 20 2 ms 100Hz 10 1 10 2 ms 50Hz 20 1 5 2 ms 25Hz ULTRASOUND APPLICATOR Piezoelectric Disk Lead zirc...

Page 31: ...t of the beam at 5mm from the transducer surface POWER AND INTENSITY Output Power 2cm2 5 cm2 1MHz 0 3W 0 12W Nominal 3MHz 0 3W 0 9W Nominal Output Intensity 1MHz 0 2W cm2 0 3W cm2 Nominal 3MHz 0 2W cm...

Page 32: ...T 2001 2017 ACCELERATED CARE PLUS CORP ALL RIGHTS RESERVED ACP ULTRASONIC THERAPY APPLICATOR INTENSITY IMAGE AT 5mm FROM TRANSDUCER SURFACE Model Number 3000C Pro Serial Number 2945 Frequency 1 or 3 M...

Page 33: ...3000C OMNISOUND 3000C PRO ULTRASOUND THERAPY SYSTEM 1 MHz and 3 MHz low BNR Ultrasound with Delta T dose control Shipping Weight 25 lbs 11 kg Standard Accessories ITEM ITEM NO DESCRIPTION 19856 AC Lin...

Page 34: ...ces Infection Control 6 x 9 perforated Sheets 1200 sheets roll 50593 Barrier Film for Surfaces Infection Control 12 x 14 perforated Sheets 800 sheets roll 63574 Barrier Tubing 3 x 004 High Clarity 1 2...

Page 35: ...ollowing have occurred 1 The product contains repairs or replacement parts not furnished by ACP 2 The product is damaged resulting from misuse or negligence 3 The product has been tampered with and or...

Page 36: ...warranty ACP is not responsible for any loss or damage to the Equipment prior to receipt at the ACP Service Center Equipment returned for warranty service must be shipped complete with all accessorie...